Home pregnancy test compared to standard-of-care ultrasound dating in the assessment of pregnancy duration.

S Johnson1, R Shaw, P Parkinson, J Ellis, P Buchanan, M Zinaman.   

Abstract

OBJECTIVE: To determine the level of agreement between the Clearblue Digital Pregnancy Test with Conception Indicator home pregnancy test and standard-of-care ultrasound in assessing pregnancy duration in a real-life, observational setting encompassing routine, clinical care. RESEARCH DESIGN AND METHODS: This was a prospective observational study of non-pregnant women seeking conception. Women collected daily urine samples from day 1 of their next menstrual cycle. If any volunteer became pregnant, daily urine samples continued to be collected for 43 days after the LH surge. Samples from day -7 to day +28 relative to the expected period (LH surge + 15 days) were tested using the home pregnancy test. This categorised any resulting pregnancies into one of three groups: 1-2 weeks, 2-3 weeks, and 3+ weeks since conception. Information from the standard UK ultrasound dating scan was also recorded by the midwife, including the expected delivery date according to ultrasound and the expected delivery date according to LMP. MAIN OUTCOME MEASURES: Full data were available from 52 pregnant women who had conceived naturally. During the study analysis, 4786 urine samples were cross-compared with 52 routine 12-week NHS ultrasound assessments and the level of agreement between home pregnancy testing and standard-of-care ultrasound in determining pregnancy duration was calculated.
RESULTS: The agreement between the gestational age as calculated by the home pregnancy test result and the exact midwife-recorded gestational age using ultrasound was 82.3%. However, when a ± 5-day range was applied to the ultrasound reading (as per routine UK clinical practice), the level of agreement was 98%.
CONCLUSIONS: The home pregnancy test provides a significantly high (98%) level of agreement with standard-of-care ultrasound when assessing pregnancy duration in a real-life, observational setting which closely mirrors daily clinical practice.

Entities:  

Mesh:

Year:  2010        PMID: 21175374     DOI: 10.1185/03007995.2010.545378

Source DB:  PubMed          Journal:  Curr Med Res Opin        ISSN: 0300-7995            Impact factor:   2.580


  3 in total

1.  Strips of Hope: Accuracy of Home Pregnancy Tests and New Developments.

Authors:  C Gnoth; S Johnson
Journal:  Geburtshilfe Frauenheilkd       Date:  2014-07       Impact factor: 2.915

2.  Estimating onset time from longitudinal and cross-sectional data with an application to estimating gestational age from longitudinal maternal anthropometry during pregnancy and neonatal anthropometry at birth.

Authors:  Ana Maria Ortega-Villa; Katherine L Grantz; Paul S Albert
Journal:  J R Stat Soc Ser A Stat Soc       Date:  2017-08-28       Impact factor: 2.483

3.  Gestational length assignment based on last menstrual period, first trimester crown-rump length, ovulation, and implantation timing.

Authors:  Amita A Mahendru; Charlotte S Wilhelm-Benartzi; Ian B Wilkinson; Carmel M McEniery; Sarah Johnson; Christoph Lees
Journal:  Arch Gynecol Obstet       Date:  2016-07-28       Impact factor: 2.344

  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.