| Literature DB >> 21170592 |
Chao Bian1, Jianhua Tong, Jizhou Sun, Hong Zhang, Qiannan Xue, Shanhong Xia.
Abstract
A field effect transistor (FET)-based immunosensor was developed for diabetes monitoring by detecting the concentrations of glycated hemoglobin (HbA1c) and hemoglobin (Hb). This immunosensor consists of a FET-based sensor chip and a disposable extended-gate electrode chip. The sensor chip was fabricated by standard CMOS process and was integrated with signal readout circuit. The disposable electrode chip, fabricated on polyester plastic board by Micro-Electro-Mechanical-Systems (MEMS) technique, was integrated with electrodes array and micro reaction pool. Biomolecules were immobilized on the electrode based on self-assembled monolayer and gold nanoparticles. Experimental results showed that the immunosensor achieved a linear response to HbA1c with the concentration from 4 to 24 μg/ml, and a linear response to Hb with the concentration from 60 to 180 μg/ml.Entities:
Mesh:
Substances:
Year: 2011 PMID: 21170592 DOI: 10.1007/s10544-010-9498-y
Source DB: PubMed Journal: Biomed Microdevices ISSN: 1387-2176 Impact factor: 2.838