Application of the two-dose-rate method for general recombination correction for liquid ionization chambers in continuous beams.

Jonas Andersson1, Heikki Tölli.   

Abstract

A method to correct for the general recombination losses for liquid ionization chambers in continuous beams has been developed. The proposed method has been derived from Greening's theory for continuous beams and is based on measuring the signal from a liquid ionization chamber and an air filled monitor ionization chamber at two different dose rates. The method has been tested with two plane parallel liquid ionization chambers in a continuous radiation x-ray beam with a tube voltage of 120 kV and with dose rates between 2 and 13 Gy min(-1). The liquids used as sensitive media in the chambers were isooctane (C(8)H(18)) and tetramethylsilane (Si(CH(3))(4)). The general recombination effect was studied using chamber polarizing voltages of 100, 300, 500, 700 and 900 V for both liquids. The relative standard deviation of the results for the collection efficiency with respect to general recombination was found to be a maximum of 0.7% for isooctane and 2.4% for tetramethylsilane. The results are in excellent agreement with Greening's theory for collection efficiencies over 90%. The measured and corrected signals from the liquid ionization chambers used in this work are in very good agreement with the air filled monitor chamber with respect to signal to dose linearity.

Entities:  

Mesh:

Substances:

Year:  2010        PMID: 21160114     DOI: 10.1088/0031-9155/56/2/001

Source DB:  PubMed          Journal:  Phys Med Biol        ISSN: 0031-9155            Impact factor:   3.609


  3 in total

1.  General ion recombination effect in a liquid ionization chamber in high-dose-rate pulsed photon and electron beams.

Authors:  Hideyuki Takei; Tomonori Isobe; Nozomi Kitamura; Yutaro Mori; Tetsuya Tomita; Daisuke Kobayashi; Satoshi Kamizawa; Tomoharu Sato; Hideyuki Sakurai; Takeji Sakae
Journal:  J Radiat Res       Date:  2018-05-01       Impact factor: 2.724

2.  Pretreatment Dose Verification in Volumetric Modulated Arc Therapy Using Liquid Ionization Chamber.

Authors:  Nitin R Kakade; Rajesh Kumar; Sunil Dutt Sharma; Vikram Mittal; D Datta
Journal:  J Med Phys       Date:  2019 Jan-Mar

3.  Characterization of recombination effects in a liquid ionization chamber used for the dosimetry of a radiosurgical accelerator.

Authors:  Antoine Wagner; Frederik Crop; Thomas Lacornerie; Nick Reynaert
Journal:  J Vis Exp       Date:  2014-05-09       Impact factor: 1.355

  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.