Dosimetric optimization of a conical breast brachytherapy applicator for improved skin dose sparing.

Yun Yang1, Mark J Rivard.   

Abstract

PURPOSE: Both the AccuBoost D-shaped and round applicators have been dosimetrically characterized and clinically used to treat patients with breast cancer. While the round applicators provide conformal dose coverage, under certain clinical circumstances the breast skin dose may be higher than preferred. The purpose of this study was to modify the round applicators to minimize skin dose while not substantially affecting dose uniformity within the target volume and reducing the treatment time.
METHODS: In order to irradiate the intended volume while sparing critical structures such as the skin, the current round applicator design has been augmented through the addition of an internal truncated cone (i.e., frustum) shield. Monte Carlo methods and clinical constraints were used to design the optimal cone applicator. With the cone applicator now defined as the entire assembly including the surrounding tungsten-alloy shell holding the HDR 192Ir source catheter, the applicator height was reduced to diminish the treatment time while minimizing skin dose. Monte Carlo simulation results were validated using both radiochromic film and ionization chamber measurements based on established techniques.
RESULTS: The optimal cone applicators diminished the maximum skin dose by 15%-32% (based on the applicator diameter and breast separation) with the tumor dose reduced by less than 3% for a constant exposure time. Furthermore, reduction in applicator height diminished the treatment time by up to 30%. Radiochromic film and ionization chamber dosimetric results in phantom agreed with Monte Carlo simulation results typically within 3%. Larger differences were outside the treatment volume in low dose regions or associated with differences between the measurement and Monte Carlo simulation environments.
CONCLUSIONS: A new radiotherapy treatment device was developed and dosimetrically characterized. This set of applicators significantly reduces the skin dose and treatment time while retaining uniform target dose.

Entities:  

Mesh:

Substances:

Year:  2010        PMID: 21158278     DOI: 10.1118/1.3495539

Source DB:  PubMed          Journal:  Med Phys        ISSN: 0094-2405            Impact factor:   4.071


  5 in total

1.  On the feasibility of polyurethane based 3D dosimeters with optical CT for dosimetric verification of low energy photon brachytherapy seeds.

Authors:  Justus Adamson; Yun Yang; Titania Juang; Kelsey Chisholm; Leith Rankine; John Adamovics; Fang Fang Yin; Mark Oldham
Journal:  Med Phys       Date:  2014-07       Impact factor: 4.071

2.  Treatment planning of a skin-sparing conical breast brachytherapy applicator using conventional brachytherapy software.

Authors:  Yun Yang; Christopher S Melhus; Shirin Sioshansi; Mark J Rivard
Journal:  Med Phys       Date:  2011-03       Impact factor: 4.071

3.  Multi-axis dose accumulation of noninvasive image-guided breast brachytherapy through biomechanical modeling of tissue deformation using the finite element method.

Authors:  Mark J Rivard; Hamid R Ghadyani; Adam D Bastien; Nicholas N Lutz; Jaroslaw T Hepel
Journal:  J Contemp Brachytherapy       Date:  2015-02-17

4.  Automated position and size selection of round applicators for AccuBoost breast brachytherapy.

Authors:  Foster L West; Reshma Munbodh; John C Patrick; Mark J Rivard; Sean A Roles; Ziad H Saleh
Journal:  J Contemp Brachytherapy       Date:  2020-12-16

5.  Optically stimulated luminescent dosimetry for high dose rate brachytherapy.

Authors:  Christopher Jason Tien; Robert Ebeling; Jessica R Hiatt; Bruce Curran; Edward Sternick
Journal:  Front Oncol       Date:  2012-08-06       Impact factor: 6.244

  5 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.