Transscleral visible/near-infrared spectroscopy for quantitative assessment of haemoglobin in experimental choroidal tumours.

Can T Xu1, Pontus Svenmarker, Stefan Andersson-Engels, Jørgen Krohn.   

Abstract

PURPOSE: To study the feasibility of using transscleral visible/near-infrared spectroscopy (Vis/NIRS) to estimate the content of haemoglobin in choroidal tumour phantoms of ex vivo porcine eyes.
METHODS: Thirty enucleated porcine eyes were prepared with a tumour phantom made by injecting a suspension of gelatine, titanium dioxide and human blood into the suprachoroidal space. The blood concentrations used were 2.5%, 25% and 50%, with 10 eyes in each group. Alternating Vis/NIRS measurements were taken over the phantom inclusion and on the opposite (normal) side of each eye. For statistical analysis, a genetic algorithm was utilized to suppress insignificant wavelengths in the spectra. The processed spectra were then used to build a regression model based on partial least squares regression and evaluated by twofold cross-validation.
RESULTS: Ultrasonography revealed that all phantoms were localized within the suprachoroidal space with no penetration through the retina. The largest mean diameters of the phantoms with 2.5%, 25% and 50% blood were 15.5, 15.2 and 15.7 mm, respectively (p > 0.05). The largest mean thicknesses were 4.5, 4.5 and 4.8 mm, respectively (p > 0.05). Statistical analysis of the spectral data showed that it was possible to correctly discriminate between the normal side and the tumour phantom side of the eyes in 99.88% of cases. The phantoms could be correctly classified according to their blood concentrations in 99.42% of cases.
CONCLUSIONS: This study demonstrates that transscleral Vis/NIRS is a feasible and accurate method for the detection of choroidal tumours and to assess the haemoglobin content in such lesions.
© 2010 The Authors. Acta Ophthalmologica © 2010 Acta Ophthalmologica Scandinavica Foundation.

Entities:  

Mesh:

Substances:

Year:  2010        PMID: 21155981     DOI: 10.1111/j.1755-3768.2010.02037.x

Source DB:  PubMed          Journal:  Acta Ophthalmol        ISSN: 1755-375X            Impact factor:   3.761


  2 in total

1.  Effects of probe geometry on transscleral diffuse optical spectroscopy.

Authors:  Pontus Svenmarker; Can T Xu; Stefan Andersson-Engels; Jørgen Krohn
Journal:  Biomed Opt Express       Date:  2011-10-07       Impact factor: 3.732

2.  Cherenkov Luminescence Imaging for Assessment of Radioactive Plaque Position in Brachytherapy of Uveal Melanoma: An In Vivo Feasibility Study.

Authors:  Jørgen Krohn; Yi-Chun Chen; Nils Ole Stabo-Eeg; Børge Hamre
Journal:  Transl Vis Sci Technol       Date:  2020-06-29       Impact factor: 3.283

  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.