Literature DB >> 21153453

Comparative study of anatomical normalization errors in SPM and 3D-SSP using digital brain phantom.

Hideo Onishi1, Yuki Matsutake, Hiroki Kawashima, Norikazu Matsutomo, Hizuru Amijima.   

Abstract

OBJECTIVE: In single photon emission computed tomography (SPECT) cerebral blood flow studies, two major algorithms are widely used statistical parametric mapping (SPM) and three-dimensional stereotactic surface projections (3D-SSP). The aim of this study is to compare an SPM algorithm-based easy Z score imaging system (eZIS) and a 3D-SSP system in the errors of anatomical standardization using 3D-digital brain phantom images.
METHODS: We developed a 3D-brain digital phantom based on MR images to simulate the effects of head tilt, perfusion defective region size, and count value reduction rate on the SPECT images. This digital phantom was used to compare the errors of anatomical standardization by the eZIS and the 3D-SSP algorithms.
RESULTS: While the eZIS allowed accurate standardization of the images of the phantom simulating a head in rotation, lateroflexion, anteflexion, or retroflexion without angle dependency, the standardization by 3D-SSP was not accurate enough at approximately 25° or more head tilt. When the simulated head contained perfusion defective regions, one of the 3D-SSP images showed an error of 6.9% from the true value. Meanwhile, one of the eZIS images showed an error as large as 63.4%, revealing a significant underestimation.
CONCLUSION: When required to evaluate regions with decreased perfusion due to such causes as hemodynamic cerebral ischemia, the 3D-SSP is desirable. In a statistical image analysis, we must reconfirm the image after anatomical standardization by all means.

Entities:  

Mesh:

Year:  2010        PMID: 21153453     DOI: 10.1007/s12149-010-0448-9

Source DB:  PubMed          Journal:  Ann Nucl Med        ISSN: 0914-7187            Impact factor:   2.668


  2 in total

1.  Influences of reconstruction and attenuation correction in brain SPECT images obtained by the hybrid SPECT/CT device: evaluation with a 3-dimensional brain phantom.

Authors:  Mana Akamatsu; Yasuo Yamashita; Go Akamatsu; Yuji Tsutsui; Nobuyoshi Ohya; Yasuhiko Nakamura; Masayuki Sasaki
Journal:  Asia Ocean J Nucl Med Biol       Date:  2014

2.  Towards characterizing the regional cerebral perfusion in evaluating the severity of major depression disorder with SPECT/CT.

Authors:  Jinming Li; Yuan Yang; Yuankai Zhu; Liqiang Zhou; Yunfeng Han; Tao Yin; Zhaoting Cheng; Guopeng Zhang; Yanxia Shen; Jing Chen
Journal:  BMC Psychiatry       Date:  2018-03-21       Impact factor: 3.630

  2 in total