Literature DB >> 21151334

Comparison of CDE data in phacoemulsification between an open hospital-based ambulatory surgical center and a free-standing ambulatory surgical center.

Ming Chen1, Mindy Chen.   

Abstract

Mean CDE (cumulative dissipated energy) values were compared for an open hospital- based surgical center and a free-standing surgical center. The same model of phacoemulsifier (Alcon Infiniti Ozil) was used. Mean CDE values showed that surgeons (individual private practice) at the free-standing surgical center were more efficient than surgeons (individual private practice) at the open hospital-based surgical center (mean CDE at the hospital-based surgical center 18.96 seconds [SD = 12.51]; mean CDE at the free-standing surgical center 13.2 seconds [SD = 9.5]). CDE can be used to monitor the efficiency of a cataract surgeon and surgical center in phacoemulsification. The CDE value may be used by institutions as one of the indicators for quality control and audit in phacoemulsification.

Entities:  

Keywords:  CDE (cumulative dissipated energy); free-standing surgical center; open hospital-based ambulatory surgical center; phacoemulsification

Year:  2010        PMID: 21151334      PMCID: PMC2993129          DOI: 10.2147/OPTH.S15076

Source DB:  PubMed          Journal:  Clin Ophthalmol        ISSN: 1177-5467


Introduction

CDE (cumulative dissipated energy) is a built-in device of the Alcon Infiniti phacoemulsification unit (Alcon Labs, Hünenberg, Switzerland). It shows the time taken in seconds to complete a case of phacoemulsification, which reflects how much energy will be expended. Less phacoemulsification time translates to less energy used in the eye, and is considered better for corneal endothelium.1–4 The purpose of this study was to introduce the CDE as a monitor or measure of surgical efficiency in the hope of improving surgical outcomes. CDE data are not currently being widely utilized for this purpose. A higher CDE reading is equated with longer surgical and recovery time, because more energy is dissipated in the eye, resulting in more damage to ocular tissue. Successful phacoemulsification of the cataract with less CDE may correlate with more efficient surgery and better overall outcomes. A previous study used CDE to compare torsional mode with conventional ultrasound mode phacoemulsification.5 Another study utilized the CDE to compare different techniques and settings by different surgeons in phacoemulsification.6

Objective

This study set out to compare mean CDE between surgeons in an open hospital- based surgical center and a free-standing surgical center, using the same model of phacoemulsifier (Alcon Infiniti Ozil). It also intended to establish the mean CDE value in the two institutions as a reference for future training and quality control in cataract surgery.

Method

Ninety cases were retrospectively selected from the records of each institution’s operating room records in one randomly selected month. Because all 180 cases were selected from one randomly selected month of the year and from similar ambulatory settings of the two nearby ambulatory surgical centers with similar descriptive data, the variety of cataract cases (soft vs hard) was expected to be similar. CDE value, taken from the phacoemulsifier, was recorded by the nurse after phacoemulsification. The CDE values of 90 cases from each institution were recorded in an Excel spreadsheet together with age and gender. There is no necessity to record the patient’s identity, which avoids privacy intrusion. The data were then analyzed by SPSS. The ratio of males and females was equal at the hospital and 45.2% versus 54.8% at the surgical center. The age distribution was from 28 to 94 years for the hospital versus 20 to 99 years for the surgical center. This study was approved by the IRB (Institutional Review Board) of the University of Hawaii for ethical issues.

Result

The mean CDE at the hospital-based center was 18.96 seconds (SD = 12.51); for the surgical center it was 13.2 seconds (SD = 9.5) (Figure 1).
Figure 1

Mean CDE in hospital-based ambulatory and free-standing centers.

Conclusion

CDE values showed that surgeons at the free-standing ambulatory surgical center were more efficient in phacoemulsification than surgeons at the open hospital-based ambulatory surgical center. The free-standing cataract surgical center may be more efficient in cataract surgery than the open hospital-based ambulatory surgical center.

Discussion

This study intended to use CDE as an objective guide to evaluate the efficiency of phacoemulsification. The open hospital-based ambulatory surgical center has more surgeons (individual private practice) on staff, and each surgeon has a different volume of cataract surgeries. The free-standing cataract surgical center only has four surgeons on staff (individual private practice), and each surgeon has a uniformly large volume of cataract surgeries. Furthermore, the majority of the free-standing surgical center’s cases are cataract surgeries, whereas the hospital-based ambulatory surgical center deals with a variety of different eye cases. Even though the hospital-based ambulatory surgical center is hospital based, it is still an ambulatory center. It is designed for ambulatory care to follow the ambulatory surgery guidelines, as with the free-standing surgical center. Because the cataract surgery was not performed in the main operation room of the hospital but rather in the ambulatory center, cataract cases selected should not be more difficult than those in the free-standing surgical center. Because all 180 cases were selected randomly from one month of the year and from similar ambulatory settings of the two nearby ambulatory surgical centers with similar descriptive data, the variety of cataract cases (soft vs hard) was expected to be similar. This study showed that the high-volume cataract surgeons at the free-standing surgical center had a low mean CDE because of their better efficiency in phacoemulsification. CDE can be used as a monitor for a cataract surgeons’ efficiency in phacoemulsification. A mean CDE value was established for the two institutions in this study. The CDE value may be used by institutions as an indicator of quality control and audit in phacoemulsification.
  6 in total

1.  Endothelial cell loss following "modern" phacoemulsification by a senior resident.

Authors:  K R Kreisler; S W Mortenson; N Mamalis
Journal:  Ophthalmic Surg       Date:  1992-03

2.  Torsional mode versus conventional ultrasound mode phacoemulsification: randomized comparative clinical study.

Authors:  Yizhi Liu; Mingbing Zeng; Xialin Liu; Lixia Luo; Zhaohui Yuan; Yuanlin Xia; Yangfa Zeng
Journal:  J Cataract Refract Surg       Date:  2007-02       Impact factor: 3.351

3.  Risk factors for corneal endothelial injury during phacoemulsification.

Authors:  K Hayashi; H Hayashi; F Nakao; F Hayashi
Journal:  J Cataract Refract Surg       Date:  1996-10       Impact factor: 3.351

4.  Effects of phacoemulsification time on the corneal endothelium using phacofracture and phaco chop techniques.

Authors:  G Pirazzoli; D D'Eliseo; M Ziosi; R Acciarri
Journal:  J Cataract Refract Surg       Date:  1996-09       Impact factor: 3.351

5.  Endothelial cell loss following phacoemulsification in the pupillary plane.

Authors:  D M Colvard; R P Kratz; T R Mazzocco; B Davidson
Journal:  J Am Intraocul Implant Soc       Date:  1981

6.  A retrospective randomized study to compare the energy delivered using CDE with different techniques and OZil settings by different surgeons in phacoemulsification.

Authors:  Ming Chen; Henry W Sweeney; Becky Luke; Mindy Chen; Mathew Brown
Journal:  Clin Ophthalmol       Date:  2009-07-14
  6 in total
  5 in total

1.  Differences in energy and corneal endothelium between femtosecond laser-assisted and conventional cataract surgeries: prospective, intraindividual, randomized controlled trial.

Authors:  Lucia Bascaran; Txomin Alberdi; Itziar Martinez-Soroa; Cristina Sarasqueta; Javier Mendicute
Journal:  Int J Ophthalmol       Date:  2018-08-18       Impact factor: 1.779

2.  Preexisting epiretinal membrane is associated with pseudophakic cystoid macular edema.

Authors:  Friederike Schaub; Werner Adler; Philip Enders; Meike C Koenig; Konrad R Koch; Claus Cursiefen; Bernd Kirchhof; Ludwig M Heindl
Journal:  Graefes Arch Clin Exp Ophthalmol       Date:  2018-03-21       Impact factor: 3.117

3.  Comparison of perioperative parameters in one-handed rotational phacoemulsification versus conventional phacoemulsification and femtosecond laser-assisted cataract surgery.

Authors:  Samuele Gigliola; Giancarlo Sborgia; Alfredo Niro; Carmela Palmisano; Pasquale Puzo; Gianluigi Giuliani; Luigi Sborgia; Dario Sisto; Valentina Pastore; Claudio Furino; Rossella Donghia; Alessandra Sborgia; Francesco Boscia; Giovanni Alessio
Journal:  Int J Ophthalmol       Date:  2021-12-18       Impact factor: 1.779

4.  Combined 30-degree bevel up and down technique against 0-degree phaco tip for phacoemulsification surgery of hard cataracts.

Authors:  Rajesh Subhash Joshi; Sonal Jayant Muley
Journal:  Clin Ophthalmol       Date:  2017-06-06

5.  Comparison of cumulative dissipated energy between the Infiniti and Centurion phacoemulsification systems.

Authors:  Ming Chen; Erik Anderson; Geoffrey Hill; John J Chen; Thomas Patrianakos
Journal:  Clin Ophthalmol       Date:  2015-07-22
  5 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.