Measurement of body temperature in adult patients: comparative study of accuracy, reliability and validity of different devices.

J Rubia-Rubia1, A Arias, A Sierra, A Aguirre-Jaime.   

Abstract

BACKGROUND AND AIMS: We compared a range of alternative devices with core body temperature measured at the pulmonary artery to identify the most valid and reliable instrument for measuring temperature in routine conditions in health services.
METHODS: 201 patients from the intensive care unit of the Candelaria University Hospital, Canary Islands, admitted to hospital between April 2006 and July 2007. All patients (or their families) gave informed consent. Readings from gallium-in-glass, reactive strip and digital in axilla, infra-red ear and frontal thermometers were compared with the pulmonary artery core temperature simultaneously. External factors suspected of having an influence on the differences were explored. The cut-off point readings for each thermometer were fixed for the maximum negative predictive value in comparison with the core temperature. The validity, reliability, accuracy, external influence, the waste they generated, ease of use, speed, durability, security, comfort and cost of each thermometer was evaluated. An ad hoc overall valuation score was obtained from these parameters for each instrument.
RESULTS: For an error of ± 0.2°C and concordance with respect to fever, the gallium-in-glass thermometer gave the best results. The largest area under the receiver operating characteristic (ROC) curve is obtained by the digital axillar thermometer with probe (0.988 ± 0.007). The minimum difference between readings was given by the infrared ear thermometer, in comparison with the core temperature (-0.1 ± 0.3°C). Age, weight, level of conscience, male sex, environmental temperature and vaso-constrictor medication increases the difference in the readings and fever treatment reduces it, although this is not the same for all thermometers. The compact digital axillar thermometer and the digital thermometer with probe obtained the highest overall valuation score.
CONCLUSION: If we only evaluate the aspects of validity, reliability, accuracy and external influence, the best thermometer would be the gallium-in-glass after 12 min. The gallium-in-glass thermometer is less accurate after only 5 min in comparison with the reading taken after being placed for 12 min. If we add the evaluation of waste production, ease-of-use, speed, durability, security, patient comfort and costs, the thermometers that obtain the highest score are the compact digital and digital with probe in right axilla.
Copyright © 2010 Elsevier Ltd. All rights reserved.

Entities:  

Mesh:

Year:  2010        PMID: 21145551     DOI: 10.1016/j.ijnurstu.2010.11.003

Source DB:  PubMed          Journal:  Int J Nurs Stud        ISSN: 0020-7489            Impact factor:   5.837


  14 in total

1.  Body Temperature in Practitioners of a Yoga Breathing Technique Considered to be Heat Generating.

Authors:  Shirley Telles; Kumar Gandharva; Ram Kumar Gupta; Sachin Kumar Sharma; Acharya Balkrishna
Journal:  Int J Yoga       Date:  2020-05-01

2.  Tympanic temperature versus temporal temperature in patients with pyrexia and chills.

Authors:  Wen-Chieh Yang; Huang-Tsung Kuo; Ching-Hsiao Lin; Kang-Hsi Wu; Yu-Jun Chang; Chun-Yu Chen; Han-Ping Wu
Journal:  Medicine (Baltimore)       Date:  2016-11       Impact factor: 1.889

3.  Assessment of axillary temperature for the evaluation of normal body temperature of healthy young adults at rest in a thermoneutral environment.

Authors:  Shuri Marui; Ayaka Misawa; Yuki Tanaka; Kei Nagashima
Journal:  J Physiol Anthropol       Date:  2017-02-22       Impact factor: 2.867

4.  Early sepsis markers in patients admitted to intensive care unit with moderate-to-severe diabetic ketoacidosis.

Authors:  Florian Blanchard; Judith Charbit; Guillaume Van der Meersch; Benjamin Popoff; Adrien Picod; Regis Cohen; Frank Chemouni; Stephane Gaudry; Helene Bihan; Yves Cohen
Journal:  Ann Intensive Care       Date:  2020-05-19       Impact factor: 6.925

Review 5.  Ga-Based Alloys in Microelectronic Interconnects: A Review.

Authors:  Shiqian Liu; Keith Sweatman; Stuart McDonald; Kazuhiro Nogita
Journal:  Materials (Basel)       Date:  2018-08-08       Impact factor: 3.623

6.  Risk factors and outcomes for prolonged versus brief fever: a prospective cohort study.

Authors:  Philippe Seguin; Antoine Roquilly; Olivier Mimoz; Pascale Le Maguet; Karim Asehnoune; Sébastien Biederman; Elsa Carise; Yannick Malledant
Journal:  Crit Care       Date:  2012-08-13       Impact factor: 9.097

7.  Accuracy of tympanic temperature measurement using an infrared tympanic membrane thermometer.

Authors:  Gasim I Gasim; Imad R Musa; Mohamed T Abdien; Ishag Adam
Journal:  BMC Res Notes       Date:  2013-05-10

Review 8.  Temperature measurements with a temporal scanner: systematic review and meta-analysis.

Authors:  Håkan Geijer; Ruzan Udumyan; Georg Lohse; Ylva Nilsagård
Journal:  BMJ Open       Date:  2016-03-31       Impact factor: 2.692

9.  Temperature measurements in trauma patients: is the ear the key to the core?

Authors:  O Uleberg; S C Eidstuen; G Vangberg; E Skogvoll
Journal:  Scand J Trauma Resusc Emerg Med       Date:  2015-11-19       Impact factor: 2.953

10.  Accuracy and precision of four common peripheral temperature measurement methods in intensive care patients.

Authors:  Simin Asadian; Alireza Khatony; Gholamreza Moradi; Alireza Abdi; Mansour Rezaei
Journal:  Med Devices (Auckl)       Date:  2016-09-01
View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.