Suitability of an automated nucleic acid extractor (easyMAG) for use with hepatitis C virus and human immunodeficiency virus type 1 nucleic acid amplification testing.

L M Jarvis1, K Mulligan, T H Dunsford, K McGowan, J Petrik.   

Abstract

Serological screening assays have greatly reduced, but not eliminated, the risk of transmission of viral infections by transfusion of blood and blood products. In addition, the 1999 regulation of the European Agency for the Evaluation of Medicinal Products requiring all plasma for fractionation to have tested negative for hepatitis C virus (HCV) RNA (CPMP/BWP/390/97, 1998) led many blood transfusion services to introduce nucleic acid amplification technology (NAT) to screen blood donations for HCV, and in some services for human immunodeficiency virus (HIV) and hepatitis B virus (HBV). BioMérieux's second-generation system, the NucliSENS easyMAG, was evaluated as a suitable platform for the automated extraction of nucleic acids for use with the existing SNBTS NAT assays. Two nucleic acid extraction protocols were examined, either lysis on the easyMAG (on board) or a 30-min pre-incubation of the sample with lysis buffer at 37 °C (off board). Off board lysis was found to extract nucleic acid more efficiently for both HCV and HIV NAT assays although the improvement was more marked with HIV. The 95% limit of detections (LODs) were 10.11 IU/ml (on board) and 7.21 IU/ml (off board) for HCV and 55.11 IU/ml (on board) and 34.13 (off board) for HIV. Using the more sensitive off board lysis, nucleic acid extraction specificity, robustness and reliability of the easyMAG were examined and over 10,000 Scottish blood donations (in 107 pools of 95 donations) were tested for HCV and HIV in parallel with the existing assay. The results indicate that the easyMAG is a suitable and flexible nucleic acid extraction system, providing high quality nucleic acids and a rapid response alternative to commercial, fully automated, approved blood screening platforms.
© 2010 Elsevier B.V. All rights reserved.

Entities:  

Mesh:

Substances:

Year:  2010        PMID: 21126541     DOI: 10.1016/j.jviromet.2010.11.021

Source DB:  PubMed          Journal:  J Virol Methods        ISSN: 0166-0934            Impact factor:   2.014


  2 in total

1.  Comparison of the new fully automated extraction platform eMAG to the MagNA PURE 96 and the well-established easyMAG for detection of common human respiratory viruses.

Authors:  Musa Hindiyeh; Orna Mor; Rakefet Pando; Batya Mannasse; Areej Kabat; Hadar Assraf-Zarfati; Ella Mendelson; Danit Sofer; Michal Mandelboim
Journal:  PLoS One       Date:  2019-02-19       Impact factor: 3.240

2.  Comparison of procedures for RNA-extraction from peripheral blood mononuclear cells.

Authors:  Antonio Rodríguez; Hans Duyvejonck; Jonas D Van Belleghem; Tessa Gryp; Leen Van Simaey; Stefan Vermeulen; Els Van Mechelen; Mario Vaneechoutte
Journal:  PLoS One       Date:  2020-02-21       Impact factor: 3.240

  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.