Measurement of melanoma thickness--comparison of two methods: ultrasound versus morphology.

Vygintas Kaikaris1, Donatas Samsanavičius, Rytis Rimdeika, Skaidra Valiukevičienė, Jurgita Makštienė, Juozas Pundzius.   

Abstract

OBJECTIVE: The aim of our study was to investigate the association between non-invasive ultrasound examination and morphologic test results in the measurement of cutaneous melanoma thickness influencing surgical treatment strategy.
METHODS: Our prospective clinical study has been conducted in the Clinic of Plastic and Reconstructive Surgery of Kaunas University of Medicine Hospital (KUMH) since January 2004 until October 2008. A total of 100 patients with a clear clinical diagnosis of stage I-II cutaneous melanoma were enrolled in this study. Melanoma depth was measured using a linear 14-MHz frequency ultrasound sensor (Toshiba Xario XG). Surgically removed fragments of skin-subcutaneous tissue were fixated using 10% formalin solution in the operating theatre, and sent to KUMH Clinic of Pathological Anatomy. The most informative sections were analysed for tumour thickness, according to Breslow, as well as tumour type, vascular and lymphatic invasion and dissemination. A comparative data analysis of melanoma thickness measured by ultrasound (T) preoperatively and histologically estimated cutaneous melanoma (CM) thickness according to Breslow (pT) using the Bland-Altman method was performed.
RESULTS: Higher mean difference of melanoma thickness (60 μm) between T and pT measurements was found when tumour thickness matched pT1 and pT2 categories. In cases of CM depth exceeding 2mm, mean difference of measurements between CM thickness determined by ultrasound and histological examination was lower (30 μm). Data regression analysis showed that correlation between T and pT measurements was lower when CM was thinner (1-2 mm) (Pearson's correlation coefficient, r: 0.283). In cases of thicker melanoma (>2 mm), strong and statistically significant (p<0.0001) correlation (r: 0.869) was observed.
CONCLUSIONS: Correlation between melanoma thickness measured using a linear 14-MHz frequency ultrasound sensor and histologically estimated melanoma thickness according to Breslow was lower if melanoma was thinner (1-2 mm). However, in cases of thicker melanoma (>2 mm), very strong correlation between measurements was observed.
Copyright © 2010 British Association of Plastic, Reconstructive and Aesthetic Surgeons. Published by Elsevier Ltd. All rights reserved.

Entities:  

Mesh:

Year:  2010        PMID: 21123126     DOI: 10.1016/j.bjps.2010.10.008

Source DB:  PubMed          Journal:  J Plast Reconstr Aesthet Surg        ISSN: 1748-6815            Impact factor:   2.740


  7 in total

1.  Sonography of the primary cutaneous melanoma: a review.

Authors:  Ximena Wortsman
Journal:  Radiol Res Pract       Date:  2012-03-01

Review 2.  High frequency ultrasound with color Doppler in dermatology.

Authors:  Elisa de Oliveira Barcaui; Antonio Carlos Pires Carvalho; Flavia Paiva Proença Lobo Lopes; Juan Piñeiro-Maceira; Carlos Baptista Barcaui
Journal:  An Bras Dermatol       Date:  2016 May-Jun       Impact factor: 1.896

3.  High-frequency ultrasound: an essential non-invasive tool for the pre-therapeutic assessment of basal cell carcinoma.

Authors:  Raluca Maria Bungărdean; Mircea Sebastian Şerbănescu; Horaţiu Alexandru Colosi; Maria Crişan
Journal:  Rom J Morphol Embryol       Date:  2021 Apr-Jun       Impact factor: 1.033

Review 4.  Usefulness of High-Frequency Ultrasonography in the Diagnosis of Melanoma: Mini Review.

Authors:  Maria Paola Belfiore; Alfonso Reginelli; Anna Russo; Gaetano Maria Russo; Maria Paola Rocco; Elvira Moscarella; Marilina Ferrante; Antonello Sica; Roberto Grassi; Salvatore Cappabianca
Journal:  Front Oncol       Date:  2021-06-11       Impact factor: 6.244

5.  One Step Surgery for Cutaneous Melanoma: "We Cannot Solve Our Problems with the Same Thinking We Used When We Created Them?"

Authors:  Georgi Tchernev
Journal:  Open Access Maced J Med Sci       Date:  2017-10-10

Review 6.  Skin melanoma imaging using ultrasonography: a literature review.

Authors:  Paweł Piłat; Adam Borzęcki; Mieczysław Jazienicki; Dorota Krasowska
Journal:  Postepy Dermatol Alergol       Date:  2018-06-18       Impact factor: 1.837

7.  Evaluation of the clinical usefulness of high-frequency ultrasonography in pre-operative evaluation of cutaneous melanoma - a prospective study.

Authors:  Paweł Piłat; Adam Borzęcki; Mieczysław Jazienicki; Agnieszka Gerkowicz; Franciszek Szubstarski; Dorota Krasowska
Journal:  Postepy Dermatol Alergol       Date:  2019-01-21       Impact factor: 1.837

  7 in total