Comparison of Two Different Blood Sample Tubes for Platelet Function Analysis with the Multiplate® System.

Ralph M Loreth1, Gudrun Klose.   

Abstract

BACKGROUND: For whole blood aggregation on the Multiplate® system, the use of thrombin inhibitor as anticoagulant is recommended. So far sample tubes containing liquid lepirudin were provided (Dynabyte, Munich, Germany). They are not sterile and have to be stored refrigerated. For better handling now also sterile tubes with dried hirudin, storable at room temperature, are also provided (Dynabyte). The aim of this study was to compare the performance of these both sample tubes on the Multiplate system. PATIENTS AND METHODS: Blood was collected from 30 patients, treated with aspirin and/or clopidogrel, in tubes, and ASPItest, ADPtest and TRAPtest were performed.
RESULTS: The correlation between both sample tubes was excellent with ASPItest (r(2) = 0.96), good with ADPtest (r(2) = 0.91) and also satisfying with TRAPtest (r(2) = 0.85).
CONCLUSION: Both sample tubes are well qualified for platelet function analysis with the Multiplate system. The new sterile sample tube with dried hirudin may be preferred for better handling.

Entities:  

Year:  2010        PMID: 21113252      PMCID: PMC2980514          DOI: 10.1159/000319785

Source DB:  PubMed          Journal:  Transfus Med Hemother        ISSN: 1660-3796            Impact factor:   3.747


  7 in total

1.  Platelet aggregatory responses to low-dose collagen are maintained in hirudin-anticoagulated whole blood for 24 h when stored at room temperature.

Authors:  P R Belcher; E W Muriithi; S P Day; D J Wheatley
Journal:  Platelets       Date:  2001-02       Impact factor: 3.862

2.  Effects of anticoagulants used during blood collection on human platelet function.

Authors:  Jane A May; Stanley Heptinstall
Journal:  Methods Mol Biol       Date:  2004

Review 3.  Platelet function analysis.

Authors:  Paul Harrison
Journal:  Blood Rev       Date:  2005-03       Impact factor: 8.250

4.  Dependence of human platelet functional responses on divalent cations: aggregation and secretion in heparin- and hirudin-anticoagulated platelet-rich plasma and the effects of chelating agents.

Authors:  B Lages; H J Weiss
Journal:  Thromb Haemost       Date:  1981-04-30       Impact factor: 5.249

5.  Multiple electrode aggregometry: a new device to measure platelet aggregation in whole blood.

Authors:  Orsolya Tóth; Andreas Calatzis; Sandra Penz; Hajna Losonczy; Wolfgang Siess
Journal:  Thromb Haemost       Date:  2006-12       Impact factor: 5.249

6.  Platelet functions in recombinant hirudin-anticoagulated blood.

Authors:  E Glusa; F Markwardt
Journal:  Haemostasis       Date:  1990

7.  The effects of citrate and extracellular calcium ions on the platelet release reaction induced by adenosine diphosphate and collagen.

Authors:  S Heptinstall; P M Taylor
Journal:  Thromb Haemost       Date:  1979-08-31       Impact factor: 5.249

  7 in total
  1 in total

1.  Clinical and Laboratory Diagnosis of Inherited Platelet Function Disorders.

Authors:  Peter Bugert
Journal:  Transfus Med Hemother       Date:  2010       Impact factor: 3.747

  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.