Hydroxyl free radical production during torsional phacoemulsification.

Steven D Aust1, Thomas Hebdon, Jordan Humbert, Ramon Dimalanta.   

Abstract

PURPOSE: To quantitate free radical generation during phacoemulsification using an ultrasonic phacoemulsification device that includes a torsional mode and evaluate tip designs specific to the torsional mode.
SETTING: Chemistry and Biochemistry Department, Utah State University, Logan, Utah, USA.
DESIGN: Experimental study.
METHODS: Experiments were performed using the Infiniti Vision System and OZil handpiece. Hydroxyl radical concentrations in the aspirated irrigation solution during torsional phacoemulsification were quantitated as nanomolar malondialdehyde (nM MDA) and determined spectrophotometrically using the deoxyribose assay.
RESULTS: The mean free radical production during phacoemulsification with torsional modality at 100% amplitude was 30.1 nM MDA ± 5.1 (SD) using a 0.9 mm 45-degree Kelman tapered ABS tip. With other tip designs intended for use with the torsional modality, free radical production was further reduced when fitted with the 0.9 mm 45-degree Kelman mini-flared ABS tip (13.2 ± 5.6 nM MDA) or the 0.9 mm 45-degree OZil-12 mini-flared ABS tip (14.3 ± 6.7 nM MDA). Although the measurements resulting from the use of the latter 2 tips were not statistically significantly different (P ≈ .25), they were different from those of the tapered tip (P<.0001).
CONCLUSIONS: The MDA concentration in the aspirated irrigation solution using the torsional modality was approximately one half that reported for the handpiece's longitudinal modality in a previous study using the same bent-tip design (Kelman tapered, P<.0001). The level of MDA was further reduced approximately one half with torsional-specific tips.
Copyright © 2010 ASCRS and ESCRS. Published by Elsevier Inc. All rights reserved.

Entities:  

Mesh:

Substances:

Year:  2010        PMID: 21111319     DOI: 10.1016/j.jcrs.2010.06.063

Source DB:  PubMed          Journal:  J Cataract Refract Surg        ISSN: 0886-3350            Impact factor:   3.351


  1 in total

1.  Effect of longitudinal and torsional ultrasound on corneal endothelial cells: experimental study in rabbit eyes.

Authors:  Emilie L Ungricht; Catherine Culp; Phillip Qu; Nathan Jensen; Jacob T Harris; Ben J Brintz; Nick Mamalis; Randall J Olson; Liliana Werner
Journal:  J Cataract Refract Surg       Date:  2022-03-01       Impact factor: 3.528

  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.