Isocentric rotational performance of the Elekta Precise Table studied using a USB-microscope.

Hans L Riis1, Sune J Zimmermann, Poul Riis.   

Abstract

The isocentric three-dimensional performance of the Elekta Precise Table was investigated. A pointer was attached to the radiation head of the accelerator and positioned at the geometric rotational axis of the head. A USB-microscope was mounted on the treatment tabletop to measure the table position relative to the pointer tip. The table performance was mapped in terms of USB-microscope images of the pointer tip at different table angles and load configurations. The USB-microscope was used as a detector to measure the pointer tip positions with a resolution down to 0.01 mm. A new elastic model of the treatment table was developed. This model describes the performance of the treatment table quite well except from some deviations due to backlash effects. Geometric and elastic features are described through six parameters. These parameters are calculated using the linear least squares fitting technique. A new method to ensure optimal positioning of the table relative to the accelerator is presented. This method cannot eliminate systematic errors completely. To eliminate systematic errors we suggest that geometric and elastic models of the table and accelerator gantry arm are incorporated in the dose planning system.

Entities:  

Mesh:

Year:  2010        PMID: 21098925     DOI: 10.1088/0031-9155/55/24/014

Source DB:  PubMed          Journal:  Phys Med Biol        ISSN: 0031-9155            Impact factor:   3.609


  1 in total

1.  Improved accuracy for noncoplanar radiotherapy: an EPID-based method for submillimeter alignment of linear accelerator table rotation with MV isocenter.

Authors:  Matthew J Nyflot; Ning Cao; Juergen Meyer; Eric C Ford
Journal:  J Appl Clin Med Phys       Date:  2014-03-06       Impact factor: 2.102

  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.