Literature DB >> 21096702

Implementation of array sensor module for a radial artery tonometry.

Ki Young Shin1, Sung Chae Jeon, Ki Chang Nam, Young Huh.   

Abstract

In this study, we developed an array sensor module to measure pulse wave of radial artery. To measure pressure, this module contains 7 piezoresistive sensors of the size of 1×1 mm. This sensor module is composed of a PCB for bonding sensor, PCB guide for wiring, silicon guide to keep its shape while PDMS coating, and PDMS layer protect fragile sensor and wire while faithfully transmitting the pressure of the patient's skin to the sensor. To test characteristic of pressure sensors in the sensor module, intra-chamber pressure was raised in 20mmHg step from 60mmHg to 220mmHg. The coefficient of determinant was r(2)>0.999. When the radial artery's pulse wave with developed sensor module was measured, pulse wave is clearly detected. In this study, developed array sensor module is appropriate to measure pulse wave.

Entities:  

Mesh:

Year:  2010        PMID: 21096702     DOI: 10.1109/IEMBS.2010.5627310

Source DB:  PubMed          Journal:  Annu Int Conf IEEE Eng Med Biol Soc        ISSN: 2375-7477


  1 in total

1.  Comparison between radial artery tonometry pulse analyzer and pulsed-Doppler echocardiography derived hemodynamic parameters in cardiac surgery patients: a pilot study.

Authors:  Nima Hatam; Ali Aljalloud; Rashad Zayat; Andreas Goetzenich; Ju-Yeon Lee; HeeJung Kang; So-Hyun Jansen-Park; Thomas Schmitz-Rode; Giulia Musetti; Heike Schnoering; Rüdiger Autschbach
Journal:  PeerJ       Date:  2017-12-06       Impact factor: 2.984

  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.