Literature DB >> 21096507

Hook effect correction & resistance-based cole fitting prior cole model-based analysis: experimental validation.

Ruben Buendia1, Fernando Seoane, Matthew Harris, Jennifer Caffarel, Roberto Gil.   

Abstract

The analysis of measurements of Electrical Bioimpedance (EBI) is on the increase for performing non-invasive assessment of health status and monitoring of pathophysiological mechanisms. EBI measurements might contain measurements artefacts that must be carefully removed prior to any further analysis. Cole model-based analysis is often selected when analysing EBI data and might lead to miss-conclusion if it is applied on data contaminated with measurement artefacts. The recently proposed Correction Function to eliminate the influence of the Hook Effect from EBI data and the fitting to the real part of the Cole model to extract the Cole parameters have been validated on experimental measurements. The obtained results confirm the feasible experimental use of these promising pre-processing tools that might improve the outcome of EBI applications using Cole model-based analysis.

Entities:  

Mesh:

Year:  2010        PMID: 21096507     DOI: 10.1109/IEMBS.2010.5627109

Source DB:  PubMed          Journal:  Annu Int Conf IEEE Eng Med Biol Soc        ISSN: 2375-7477


  1 in total

1.  Preliminary study on parameterization of raw electrical bioimpedance data with 3 frequencies.

Authors:  C A González-Correa; S A Jaimes; J I Cárdenas-Jiménez
Journal:  Sci Rep       Date:  2022-06-03       Impact factor: 4.996

  1 in total