Hiroto Sugano1, Tetsuo Tsujioka, Tadayuki Inoue, Shigeyoshi Nakajima, Shinsuke Hara, Hajime Nakamura, Kazuhide Takeuchi.
Abstract
This paper introduces the concept of an online medical diagnosis system for ubiquitous health care using a wireless ECG sensor. To confirm the feasibility ofthe system, we conducted clinical tests by 67 subjects with a wireless ECG sensor and a Holter ECG monitor simultaneously for comparison purpose. We made five types of evaluations such as analyses on data loss rate, burst data loss length, ECG waveforms comparison, normalized cross-correlation and heart rate variability (HRV) by RR50. The results show that, as long as the sensed data are successfully received at a receiver, the wireless ECG sensor has a comparable performance with the Holter ECG monitor.Mesh:
Year: 2010 PMID: 21096427 DOI: 10.1109/IEMBS.2010.5626793
Source DB: PubMed Journal: Annu Int Conf IEEE Eng Med Biol Soc ISSN: 2375-7477