| Literature DB >> 21096387 |
Martin Schuettler1, Juan S Ordonez, Tomas Silva Santisteban, Andreas Schatz, Juergen Wilde, Thomas Stieglitz.
Abstract
A fabrication technology for small hermetic implant packages with large numbers of electrical feedthroughs is presented. First prototypes were fabricated on a ceramic substrate of 25·25 mm area, having a metal cup of 5 mm height soldered to it. These prototypes provide 360 feedthroughs. The electrical properties of the feedthroughs are characterized and the leakage rate of the packages is determined using helium fine leak tests. The amount of water sealed inside the packages is investigated. Based on maximum allowable water vapour concentrations inside hermetic packages reported in literature and applying a commonly accepted mathematical model, we predicted a minimum lifetime to water-induced failure of a few hundred years.Entities:
Mesh:
Year: 2010 PMID: 21096387 DOI: 10.1109/IEMBS.2010.5626677
Source DB: PubMed Journal: Annu Int Conf IEEE Eng Med Biol Soc ISSN: 2375-7477