Literature DB >> 21095807

Self-powered instrumented knee implant for early detection of postoperative complications.

Shaban Almouahed1, Manuel Gouriou, Chafiaa Hamitouche, Eric Stindel, Christian Roux.   

Abstract

In-vivo measurement of tibiofemoral forces transmitted through Total Knee Replacement (TKR) during normal walking allows the early detection of postoperative complications such as the tibiofemoral misalignment and soft-tissue imbalance. In addition, the in-vivo data can help to improve the design of TKR in order to reduce polyethylene wear and consequently to increase the lifespan of knee implant. A self-powered custom-designed tibial implant instrumented with four piezoceramics has been developed in order to detect the aforementioned complications by measuring the relative change in pressure center (COP) position for different levels of eccentric compressive loading. Moreover, the energy harvested by the piezoceramics can be used to power a transmission system located at the stem of knee implant to wirelessly transmit the in-vivo data outside the implant for further processing and display.

Entities:  

Mesh:

Year:  2010        PMID: 21095807     DOI: 10.1109/IEMBS.2010.5626199

Source DB:  PubMed          Journal:  Annu Int Conf IEEE Eng Med Biol Soc        ISSN: 2375-7477


  5 in total

1.  Self-Powered Load Sensing Circuitry for Total Knee Replacement.

Authors:  Manav Jain; Nabid Aunjum Hossain; Shahrzad Towfighian; Ryan Willing; Milutin Stanaćević; Emre Salman
Journal:  IEEE Sens J       Date:  2021-09-03       Impact factor: 4.325

2.  Subcutaneous Photovoltaic Infrared Energy Harvesting for Bio-Implantable Devices.

Authors:  Eunseong Moon; David Blaauw; Jamie D Phillips
Journal:  IEEE Trans Electron Devices       Date:  2017-03-27       Impact factor: 2.917

Review 3.  Energy harvesting for the implantable biomedical devices: issues and challenges.

Authors:  Mahammad A Hannan; Saad Mutashar; Salina A Samad; Aini Hussain
Journal:  Biomed Eng Online       Date:  2014-06-20       Impact factor: 2.819

Review 4.  Power Approaches for Implantable Medical Devices.

Authors:  Achraf Ben Amar; Ammar B Kouki; Hung Cao
Journal:  Sensors (Basel)       Date:  2015-11-13       Impact factor: 3.576

5.  Performance of a Piezoelectric Energy Harvesting System for an Energy-Autonomous Instrumented Total Hip Replacement: Experimental and Numerical Evaluation.

Authors:  Hans-E Lange; Nils Arbeiter; Rainer Bader; Daniel Kluess
Journal:  Materials (Basel)       Date:  2021-09-08       Impact factor: 3.623

  5 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.