| Literature DB >> 21093247 |
Tetsuhiro Harimoto1, Kuniharu Takei, Takeshi Kawano, Akito Ishihara, Takahiro Kawashima, Hidekazu Kaneko, Makoto Ishida, Shiro Usui.
Abstract
In order to record multi-site electroretinogram (ERG) responses in isolated carp retinae, we utilized three-dimensional (3D), extracellular, 3.5-μm-diameter silicon (Si) probe arrays fabricated by the selective vapor-liquid-solid (VLS) growth method. Neural recordings with the Si microprobe exhibit low signal-to-noise (S/N) ratios of recorded responses due to the high-electrical-impedance characteristics of the small recording area at the probe tip. To increase the S/N ratio, we designed and fabricated enlarged gold (Au) tipped Si microprobes (10-μm-diameter Au tip and 3.5-μm-diameter probe body). In addition, we demonstrated that the signal attenuation and phase delay of ERG responses recorded via the Si probe can be compensated by the inverse filtering method. We conclude that the reduction of probe impedance and the compensation of recorded signals are useful approaches to obtain distortion-free recording of neural signals with high-impedance microelectrodes. Copyright ÂEntities:
Mesh:
Substances:
Year: 2010 PMID: 21093247 DOI: 10.1016/j.bios.2010.10.014
Source DB: PubMed Journal: Biosens Bioelectron ISSN: 0956-5663 Impact factor: 10.618