| Literature DB >> 21092331 |
Chih-Ping Chen1, Wen-Li Liao, Yi-Li Tseng, Pen-Li Lu, Yu-Chun Lo, You-Yin Chen, Fu-Shan Jaw.
Abstract
BACKGROUND: A mechanical noxious stimulator is useful for studies of pain, both for clinic and basic research. We propose to use a pinch stimulator that can not only generate a quantitative, reproducible noxious pinch but also simultaneously provide a synchronous external trigger signal, which is essential for acquisition of evoked potentials.Entities:
Mesh:
Year: 2010 PMID: 21092331 PMCID: PMC3009631 DOI: 10.1186/1475-925X-9-79
Source DB: PubMed Journal: Biomed Eng Online ISSN: 1475-925X Impact factor: 2.819
Figure 1The complete circuit diagram of the pinch stimulator. The three outputs include (A) the analog voltage signal that encodes the force of a pinch, (B) an alarm signal, and (C) a TTL-compatible synchronous output.
Figure 2Specifications of the pinch stimulator. (A) This photograph shows how the strain gauge attaches to the forceps (Dumont, type 2A). (B) Shown is the frequency response. (C) The transfer function of the stimulator, i.e., the relationship between the output voltage and the force produced by the forceps, is shown.
Figure 3Noxious-pinch evoked APs (× 2500). (A) To show the reproducibility of the applied forces at the pain threshold, three trials are superimposed. The starting point of the upper traces illustrates that the threshold has been reached, and a synchronous output for triggering data acquisition is simultaneously generated, which corresponds to the 0-ms time point on the x-axis. The lower portion of this panel shows the three superimposed evoked responses. (B) Comparison between APs generated by noxious pinch (upper traces) or control conditions (lower traces). The coordinates of the recorded unit was 1.85 mm lateral to the midline and 1.8 mm caudal to bregma, with a depth of 1050 μm. As shown previously [3], the fast nociceptive responses recorded from the primary somatosensory cortex had conduction latencies of approximately 60 ms (A-δ fibers). Thus, only the evoked APs that occurred during the 50 to 80 ms after pinch are shown.