Tarek Halabi1, Bruce Faddegon.
Abstract
We design a practical procedure for measuring translational and rotational misalignment of graticule with collimator axis of rotation and collimator jaws, respectively. The procedure's quantitative results are accurate to less than 0.2 mm (at isocenter) and do not assume alignment of radiation focal spot with collimator axis of rotation. When provided with these quantitative results, the manufacturer can custom-adjust graticules to the purchaser's collimator head.Entities:
Mesh:
Year: 2010 PMID: 21081892 PMCID: PMC5720415 DOI: 10.1120/jacmp.v11i4.3318
Source DB: PubMed Journal: J Appl Clin Med Phys ISSN: 1526-9914 Impact factor: 2.102
Figure 1Double exposure film of graticule.
Figure 2Shifts obtained from line fits to digitized graticule dots of film. Shifts apply at 0° collimator angle.