Noise exposure is increased with neonatal helmet CPAP in comparison with conventional nasal CPAP.

D Trevisanuto1, L Camiletti, N Doglioni, F Cavallin, A Udilano, V Zanardo.   

Abstract

BACKGROUND: in adults, noninvasive ventilation via a helmet is associated with significantly greater noise than nasal and facial masks. We hypothesized that noise exposure could be increased with neonatal helmet continuous positive airway pressure (CPAP) in comparison with conventional nasal CPAP (nCPAP). Our primary objective was to compare the noise intensity produced by a neonatal helmet CPAP and a conventional nCPAP system. Furthermore, we aimed to evaluate the effect of the gas flow rate and the presence of the humidifier and the filter on noise levels during neonatal helmet CPAP treatment.
METHODS: in this bench study, noise intensity was measured in the following settings: helmet CPAP, nCPAP, incubator and the neonatal intensive care unit. In helmet CPAP, noise measurements were performed at different gas flow rates (8, 10 and 12 l/min), while in nCPAP, the flow rate was 8 l/min. For both CPAP systems, the level of pressure was maintained constant at 5 cmH(2) O.
RESULTS: during neonatal helmet CPAP, the median (interquartile range) noise levels were significantly higher than those during nCPAP: 70.0 dB (69.9-70.4) vs. 62.7 dB (62.5-63.0); P<0.001. In the helmet CPAP, the noise intensities changed with increasing flow rate and with the presence of a humidifier or a filter.
CONCLUSIONS: noise intensities generated by the neonatal helmet CPAP were significantly higher than those registered while using a conventional nCPAP system. In the helmet, the noise intensity depends on the gas flow rate, and the presence of a humidifier and a filter in the system. 2010 The Acta Anaesthesiologica Scandinavica Foundation.

Mesh:

Year:  2010        PMID: 21077846     DOI: 10.1111/j.1399-6576.2010.02356.x

Source DB:  PubMed          Journal:  Acta Anaesthesiol Scand        ISSN: 0001-5172            Impact factor:   2.105


  6 in total

Review 1.  CPAP review.

Authors:  Olie Chowdhury; Catherine J Wedderburn; Donovan Duffy; Anne Greenough
Journal:  Eur J Pediatr       Date:  2011-12-16       Impact factor: 3.183

2.  Sound reduction management in the neonatal intensive care unit for preterm or very low birth weight infants.

Authors:  Abdulraoof Almadhoob; Arne Ohlsson
Journal:  Cochrane Database Syst Rev       Date:  2020-01-27

3.  Effect of Filters on the Noise Generated by Continuous Positive Airway Pressure Delivered via a Helmet.

Authors:  Ricardo Hernández-Molina; Francisco Fernández-Zacarías; Isabel Benavente-Fernández; Gema Jiménez-Gómez; Simón Lubián-López
Journal:  Noise Health       Date:  2017 Jan-Feb       Impact factor: 0.867

4.  Investigating Noise Exposure to Newborn Infants From Respiratory Support: Methodological Considerations.

Authors:  Devika Singh; Gerhard Fusch
Journal:  Cureus       Date:  2021-11-08

Review 5.  The use of head helmets to deliver noninvasive ventilatory support: a comprehensive review of technical aspects and clinical findings.

Authors:  Andrea Coppadoro; Elisabetta Zago; Fabio Pavan; Giuseppe Foti; Giacomo Bellani
Journal:  Crit Care       Date:  2021-09-08       Impact factor: 9.097

Review 6.  Helmet Ventilation for Pediatric Patients During the COVID-19 Pandemic: A Narrative Review.

Authors:  Shu-Chi Mu; Yu-Hsuan Chien; Pin-Zhen Lai; Ke-Yun Chao
Journal:  Front Pediatr       Date:  2022-02-02       Impact factor: 3.418

  6 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.