Literature DB >> 21073149

MALDI ion trap mass spectrometer with charge detector for large biomolecule detection.

Chien-Hsun Chen1, Jung-Lee Lin, Ming-Lee Chu, Chung-Hsuan Chen.   

Abstract

Up to now, all commercial matrix-assisted laser desorption/ionization (MALDI) mass spectrometers still can not efficiently analyze very large biomolecules. In this work, we report the development of a novel MALDI ion trap mass spectrometer which can enrich biomolecular ions to enhance the detection sensitivity. A charge detector was installed to measure the large ions directly. With this design, we report the first measurement of IgM with the mass-to-charge ratio (m/z) at 980 000. In addition, quantitative measurements of the number of ions can be obtained. A step function frequency scan was first developed to get a clear signal in the m/z range from 200,000 to 1,000,000.

Entities:  

Mesh:

Substances:

Year:  2010        PMID: 21073149     DOI: 10.1021/ac102217e

Source DB:  PubMed          Journal:  Anal Chem        ISSN: 0003-2700            Impact factor:   6.986


  3 in total

1.  High Mass Ion Detection with Charge Detector Coupled to Rectilinear Ion Trap Mass Spectrometer.

Authors:  Avinash A Patil; Szu-Wei Chou; Pei-Yu Chang; Chen-Wei Lee; Chun-Yen Cheng; Ming-Lee Chu; Wen-Ping Peng
Journal:  J Am Soc Mass Spectrom       Date:  2016-12-13       Impact factor: 3.109

Review 2.  A How to Guide: Clinical Population Test Development and Authorization of MALDI-ToF Mass Spectrometry-Based Screening Tests for Viral Infections.

Authors:  Ray K Iles; Jason K Iles; Raminta Zmuidinaite; Michael Roberts
Journal:  Viruses       Date:  2022-09-03       Impact factor: 5.818

3.  Biomolecular Clusters Distribution up to Mega Dalton Region Using MALDI-Quadrupole Ion Trap Mass Spectrometer.

Authors:  Yung-Kun Chuang; Szu-Hsueh Lai; Jung-Lee Lin; Chung-Hsuan Chen
Journal:  Int J Mol Sci       Date:  2018-09-17       Impact factor: 5.923