Abstract
OBJECTIVE: This study aimed to compare gas exchange with heliox and oxygen-enriched air during piston-type high-frequency oscillatory ventilation (HFOV). We hypothesized that helium gas would improve both carbon dioxide elimination and arterial oxygenation during piston-type HFOV.Entities:
Mesh:
Substances:
Year: 2010 PMID: 21070673 PMCID: PMC2993714 DOI: 10.1186/1475-925X-9-71
Source DB: PubMed Journal: Biomed Eng Online ISSN: 1475-925X Impact factor: 2.819
MAP, AMP, PaO2 and PaCO2 values during the experiment, FiO2, 0.5; Frequency, 15 Hz, Values represent mean ± SD. * P < 0.05,
| Experimental series | 1 | 2 | 3 | 4 | 5 |
|---|---|---|---|---|---|
| MAP (cmH2O) | 14 ± 0.55 | 11.5 ± 0.72* | 13.8 ± 0.6 | 11.9 ± 0.85* | 13.8 ± 0.88 |
| AMP (cmH2O) | 40.1 ± 0.61 | 28.7 ± 0.51* | 39.8 ± 1.22 | 38.4 ± 0.95 | 39.8 ± 0.61 |
| PaO2 (mmHg) | 207 ± 8.6 | 196 ± 15.5 | 210 ± 26.8 | 204 ± 18.4 | 214 ± 24.3 |
| PaCO2 (mmHg) | 43 ± 13 | 47 ± 14 | 45 ± 11 | 33 ± 5* | 43 ± 6 |