Literature DB >> 21063000

Incidence and management of CoreValve dislocation during transcatheter aortic valve implantation.

Sarah Geisbüsch1, Sabine Bleiziffer, Domenico Mazzitelli, Hendrik Ruge, Robert Bauernschmitt, Rüdiger Lange.   

Abstract

BACKGROUND: Transcatheter aortic valve implantation is a highly specialized technique offering a new therapeutic option to patients at high risk for conventional surgery. Complications associated with this catheter procedure differ from complications after surgical aortic valve replacement. This is to report incidence, management, and impact on morbidity and mortality of CoreValve dislocation during catheter valve implantation. METHODS AND
RESULTS: Between June 2007 and September 2009, the self-expandable CoreValve prosthesis (Medtronic Inc, Minneapolis, Minn) was implanted in 212 patients through a transarterial (femoral or subclavian) access. Patients with severe aortic stenosis who were at high risk for conventional surgery were included. We observed dislocation of the prosthesis during CoreValve implantation in 21 patients. In 16 cases, the CoreValve could be implanted in the correct annular position after retrieving it and reloading the catheter. In 4 patients, the completely deployed prosthesis had to be placed in the ascending or abdominal aorta before implanting a second one. One patient underwent open surgery. Overall 30-day mortality was 11.0%, 21.5% in the dislocation group and 9.9% in patients without dislocation (P=0.024). Coronary ischemia, stroke, and renal failure occurred more frequently in patients with dislocation, whereas pacemaker dependency did not differ significantly between the groups.
CONCLUSIONS: CoreValve dislocation during transcatheter aortic valve implantation occurred in 10% of the cases and significantly increases perioperative risk for severe complications or death. It requires individual specific management and can be managed either interventionally or, rarely, results in open surgery.

Entities:  

Mesh:

Year:  2010        PMID: 21063000     DOI: 10.1161/CIRCINTERVENTIONS.110.944983

Source DB:  PubMed          Journal:  Circ Cardiovasc Interv        ISSN: 1941-7640            Impact factor:   6.546


  7 in total

1.  Robotic-assisted real-time MRI-guided TAVR: from system deployment to in vivo experiment in swine model.

Authors:  Joshua L Chan; Dumitru Mazilu; Justin G Miller; Timothy Hunt; Keith A Horvath; Ming Li
Journal:  Int J Comput Assist Radiol Surg       Date:  2016-05-31       Impact factor: 2.924

2.  Towards real-time cardiovascular magnetic resonance guided transarterial CoreValve implantation: in vivo evaluation in swine.

Authors:  Philipp Kahlert; Nina Parohl; Juliane Albert; Lena Schäfer; Renate Reinhardt; Gernot M Kaiser; Ian McDougall; Brad Decker; Björn Plicht; Raimund Erbel; Holger Eggebrecht; Mark E Ladd; Harald H Quick
Journal:  J Cardiovasc Magn Reson       Date:  2012-03-27       Impact factor: 5.364

3.  Numerical evaluation of transcatheter aortic valve performance during heart beating and its post-deployment fluid-structure interaction analysis.

Authors:  Ram P Ghosh; Gil Marom; Matteo Bianchi; Karl D'souza; Wojtek Zietak; Danny Bluestein
Journal:  Biomech Model Mechanobiol       Date:  2020-02-24

4.  The role of computed tomography in pre-procedural planning of cardiovascular surgery and intervention.

Authors:  Prabhakar Rajiah; Paul Schoenhagen
Journal:  Insights Imaging       Date:  2013-08-02

5.  The safety of introducing a new generation TAVR device: one departments experience from introducing a second generation repositionable TAVR.

Authors:  Henrik Bjursten; Shahab Nozohoor; Malin Johansson; Igor Zindovic; Carl-Fredrik Appel; Johan Sjögren; Magnus Dencker; Göran Olivecrona; Jan Harnek; Sasha Koul; Ted Feldman; Michael J Reardon; Matthias Götberg
Journal:  BMC Cardiovasc Disord       Date:  2017-01-13       Impact factor: 2.298

6.  Image quality and contrast agent exposure in cardiac computed tomography angiography prior to transcatheter aortic valve implantation procedures using different acquisition protocols.

Authors:  Nina P Hofmann; Moritz Schuetz; Raffi Bekeredjian; Sven Pleger; Emanuel Chorianopoulos; Sorin Giusca; Florian André; Gitsios Gitsioudis; Christopher Schlett; Hans-Ulrich Kauczor; Hugo A Katus; Grigorios Korosoglou
Journal:  Eur J Radiol Open       Date:  2017-06-26

7.  Micro-dislodgement during transcatheter aortic valve implantation with a contemporary self-expandable prosthesis.

Authors:  Katharina Hellhammer; Kerstin Piayda; Shazia Afzal; Verena Veulemans; Inga Hennig; Matthias Makosch; Amin Polzin; Malte Kelm; Tobias Zeus
Journal:  PLoS One       Date:  2019-11-07       Impact factor: 3.240