C H Sibata1, H C Mota, P D Higgins, D Gaisser, J P Saxton, K H Shin.
Abstract
Radiotherapy treatment of patients having a hip prosthesis is a common problem facing dosimetrists and physicists when the treatment plan requires irradiation of the pelvic area. To quantify the perturbation of these devices, attenuation studies were done with 6 and 18 MV photon beams using various hip prostheses models with varying size and composition. These studies have shown that an attenuation of as much as 50% can be found in a single beam profile under the prosthesis. We have studied the capability of a dose planning system to predict the transmission of these devices as compared with measurements.Entities:
Mesh:
Substances:
Year: 1990 PMID: 2105924 DOI: 10.1016/0360-3016(90)90115-z
Source DB: PubMed Journal: Int J Radiat Oncol Biol Phys ISSN: 0360-3016 Impact factor: 7.038