Beam characteristics of megavoltage beams at low monitor unit settings.

G Kemikler1, H Acun.   

Abstract

Beam characteristics of a linear accelerator are of great importance for intensity-modulated radiation therapy (IMRT) to ensure precise and accurate dose delivery to patients. In step-and-shoot IMRT, each beam is delivered through a series of small, segmented fields at low monitor unit (MU) settings. In this study, the beam characteristics of both static (ST) and segmental intensity-modulated (IM) beams were investigated at various dose rates for 6 and 18 MV at low MU settings. Dose linearity was investigated for both the ST and the IM beams. For the ST beams, standard 10 × 10 cm(2) fields were irradiated with MU values ranging from 1 to 100. For the IM beams, 10 × 10 cm(2) and 15 × 15 cm(2) fields were used as subfields. The normalized dose (ND)/MU was obtained. Beam flatness and symmetry for 2 and 10 MU was measured by in-plane (G-T) and cross-plane (R-L) profiles using Kodak XV films. The largest dose/MU discrepancies were observed for 1 MU. For the ST beams, the beam output decreased up to 4.5% for 1 MU at the high dose rates of 6 and 18 MV. Dose variations were less than 1% for doses above 5 MU. No significant variation was observed in the beam profiles of the ST and the IM groups. Beam flatness and symmetry were close to 3% and 2% for 6 and 18 MV, respectively. Our results showed that dose linearity and delivery errors were close to 1% for doses above 5 MU, which is considered acceptable for both 6- and 18-MV ST and IM therapy.
Copyright © 2010 Associazione Italiana di Fisica Medica. Published by Elsevier Ltd. All rights reserved.

Entities:  

Mesh:

Year:  2010        PMID: 21051256     DOI: 10.1016/j.ejmp.2010.10.007

Source DB:  PubMed          Journal:  Phys Med        ISSN: 1120-1797            Impact factor:   2.685


  3 in total

1.  Analyzing the characteristics of 6 MV photon beam at low monitor unit settings.

Authors:  L Nithya; N Arunai Nambi Raj; Sasikumar Rathinamuthu
Journal:  J Med Phys       Date:  2016 Jan-Mar

2.  The Impact of Dose Rate on the Accuracy of Step-and-Shoot Intensity-modulated Radiation Therapy Quality Assurance Using Varian 2300CD.

Authors:  Christopher F Njeh; Howard W Salmon; Claire Schiller
Journal:  J Med Phys       Date:  2017 Oct-Dec

3.  Dosimetric Comparison of Robotic and Linear Accelerator Multi-Leaf Collimator-Based Stereotactic Radiosurgery for Arteriovenous Malformation.

Authors:  Venkatesan Kaliyaperumal; Susan Abraham; Maragatha Veni; Susovan Banerjee; S Tamilselvan; Deepak Gupta; K Dayanithi; D Manigandan; Saumyaranjan Mishra; Shyam Singh Bisht; Tejinder Kataria
Journal:  J Med Phys       Date:  2021-05-05
  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.