Variability in the calibration error of the goldmann applanation tonometer.

Nikhil Shreeram Choudhari1, Vaishali Jadhav, Ronnie George, Lingam Vijaya.   

Abstract

PURPOSE: To investigate the variability in the measured calibration error with continued usage of Goldmann applanation tonometers (GATs) having unacceptable calibration error.
METHODS: The study included 132 slit-lamp mounted Goldmann tonometers (Model AT 900 C/M; Haag-Streit, Switzerland). A single observer twice checked a randomly selected set of 25 instruments on 2 consecutive days to determine the intraobserver agreement in the measurement of GAT calibration error. The same observer prospectively checked all the instruments between 8 and 9 AM on any given day at all testing levels namely 0, 20, and 60 mm Hg and rechecked the faulty instruments (calibration error more than ± 2 mm Hg at any testing level) 2 times more on the same day between 12 noon and 1 PM and 4 and 5 PM.
RESULTS: The single measures intraclass correlation coefficients for the intraobserver agreement at the 20 mm Hg testing level were 0.78 for positive error and 0.83 for negative error. Twenty-eight (21%) instruments were faulty at any testing level. Nineteen (14%) were faulty at the clinically most important 20 mm Hg testing level. The maximum observed variability in the positive and negative calibration error at any testing level was +4 and -23 mm Hg, respectively. Fifteen (53%) faulty instruments had high variability (≥ 2 mm Hg) in the calibration error at any testing level.
CONCLUSIONS: The calibration error of faulty GATs can frequently have a high variability. One should avoid estimating the true intraocular pressure from a faulty GAT by instinct.

Entities:  

Mesh:

Year:  2011        PMID: 21048512     DOI: 10.1097/IJG.0b013e3181f464b8

Source DB:  PubMed          Journal:  J Glaucoma        ISSN: 1057-0829            Impact factor:   2.503


  5 in total

1.  Potential effects of systematic errors in intraocular pressure measurements on screening for ocular hypertension.

Authors:  M J Turner; S L Graham; A P Avolio; P Mitchell
Journal:  Eye (Lond)       Date:  2013-02-22       Impact factor: 3.775

2.  Rectifying calibration error of Goldmann applanation tonometer is easy!

Authors:  Nikhil S Choudhari; Krishna P Moorthy; Vinod B Tungikar; Mohan Kumar; Ronnie George; Harsha L Rao; Sirisha Senthil; Lingam Vijaya; Chandra Sekhar Garudadri
Journal:  Indian J Ophthalmol       Date:  2014-11       Impact factor: 1.848

3.  Innovative ophthalmology.

Authors:  Sundaram Natarajan
Journal:  Indian J Ophthalmol       Date:  2014-11       Impact factor: 1.848

Review 4.  Tonometry and Care of Tonometers.

Authors:  Rajat Maheshwari; Nikhil S Choudhari; Manav Deep Singh
Journal:  J Curr Glaucoma Pract       Date:  2012-10-16

5.  Agreement and repeatability of Icare ic100 tonometer.

Authors:  Judy Jose; Ramesh S Ve; H Vijaya Pai; Sayantan Biswas; Vamsi Parimi; Paresh Poojary; T Nagarajan
Journal:  Indian J Ophthalmol       Date:  2020-10       Impact factor: 1.848

  5 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.