Mesh covered stent in ST-segment elevation myocardial infarction.

Dariusz Dudek1, Artur Dziewierz, Łukasz Rzeszutko, Jacek Legutko, Wojciech Dobrowolski, Tomasz Rakowski, Stanisław Bartus, Jacek Dragan, Artur Klecha, Alexandra-J Lansky, Zbigniew Siudak, Krzysztof Zmudka.   

Abstract

AIMS: The MGuard stent (bare metal stent wrapped externally with a polymer mesh sleeve) is designed to prevent distal embolisation by reducing thrombus and plaque fragments released during and post percutaneous coronary intervention (PCI). The aim of this study was to confirm the clinical feasibility, safety and performance of the MGuard stent during primary PCI for ST-segment elevation myocardial infarction (STEMI). METHODS AND
RESULTS: The present study was a multicentre, prospective, single arm study in which 60 patients with STEMI <12 hours were enrolled. Predilatation was performed in 61.7% of the cases and thrombus aspiration in 18.3%. In one (1.7%) patient the stent could not cross the lesion. Final TIMI grade 3 flow was observed in 90.0% of patients, with myocardial blush grade 3 in 73.3% of patients and complete (>70%) ST-segment resolution 60 minutes after PCI in 61.4% of patients. In 5.0% of cases distal embolisation occurred. The total major adverse cardiac events rate during the 6-month follow-up was 1.7%.
CONCLUSIONS: Based on this experience, the MGuard stent implantation in STEMI patients is safe and highly effective. A larger randomised trial is warranted to confirm the clinical endpoints.

Entities:  

Mesh:

Year:  2010        PMID: 21044911     DOI: 10.4244/EIJV6I5A98

Source DB:  PubMed          Journal:  EuroIntervention        ISSN: 1774-024X            Impact factor:   6.534


  5 in total

1.  MGuard Embolic Protection Stent - The Importance of Thrombus Management in ST-elevation Myocardial Infarction Primary Percutaneous Coronary Intervention: Proceedings of STEMI symposium at EuroPCR on 20-23rd May 2014 in Paris.

Authors:  Katrina Mountfort; Ricardo Costa; Alexandre Abizaid; Jean Fajadet; Chaim Lotan; Ran Kornowski; Dariusz Dudek; Jose Ps Henriques; Giovanni Amoroso
Journal:  Interv Cardiol       Date:  2014-08

Review 2.  Coronary thrombus in patients undergoing primary PCI for STEMI: Prognostic significance and management.

Authors:  Sabine Vecchio; Elisabetta Varani; Tania Chechi; Marco Balducelli; Giuseppe Vecchi; Matteo Aquilina; Giulia Ricci Lucchi; Alessandro Dal Monte; Massimo Margheri
Journal:  World J Cardiol       Date:  2014-06-26

Review 3.  The MGuard coronary stent: safety, efficacy, and clinical utility.

Authors:  Montserrat Gracida; Rafael Romaguera; Francisco Jacobi; Joan A Gómez-Hospital; Angel Cequier
Journal:  Vasc Health Risk Manag       Date:  2015-09-18

Review 4.  Pharmaco-mechanic antithrombotic strategies to reperfusion of the infarct-related artery in patients with ST-elevation acute myocardial infarctions.

Authors:  Petr Kala; Roman Miklik
Journal:  J Cardiovasc Transl Res       Date:  2013-02-14       Impact factor: 4.132

Review 5.  No reflow phenomenon in percutaneous coronary interventions in ST-segment elevation myocardial infarction.

Authors:  Sanjiv Gupta; Madan Mohan Gupta
Journal:  Indian Heart J       Date:  2016-04-19
  5 in total