| Literature DB >> 21042488 |
Hr Joshi1, Ah Patel, Ad Captain.
Abstract
Two methods are described for determination of Doxophylline in a solid dosage form. The first method was based on ultraviolet (UV)-spectrophotometric determination of the drug. It involves absorbance measurement at 274 nm (λ(max) of Doxophylline) in 0.1 N hydrochloric acid. The calibration curve was linear, with the correlation coefficient between 0.99 and 1.0 over a concentration range of 0.20-30 mg/ml for the drug. The second method was based on high-performance liquid chromatography (HPLC) separation of the drug in reverse-phase mode using the Hypersil ODS C(18) column (250 × 4.6 mm, 5 mm). The mobile phase constituted of buffer acetonitrile (80:20) and pH adjusted to 3.0, with dilute orthophosphoric acid delivered at a flow rate 1.0 ml/min. Detection was performed at 210 nm. Separation was completed within 7 min. The calibration curve was linear, with the correlation coefficient between 0.99 and 1.0 over a concentration range of 0.165-30 mg/ml for the drug. The relative standard deviation was found to be <2.0% for the UV-spectrophotometry and HPLC methods. Both these methods have been successively applied to the solid dosage pharmaceutical formulation, and were fully validated according to ICH guidelines.Entities:
Keywords: Doxophylline; HPLC; UV-spectrophotometry; reversed-phase
Year: 2010 PMID: 21042488 PMCID: PMC2964760 DOI: 10.4103/0975-1483.66791
Source DB: PubMed Journal: J Young Pharm ISSN: 0975-1483
Figure 1Structure of Doxophylline
System suitability and system precision (for HPLC)
| Compound | Retention time (Mean ± SEM) | T | k’ | |
|---|---|---|---|---|
| Doxophylline | 6.434 ± 0.06217 | 11034.808 | 1.22 | 642.4 |
n = Theoretical plates
T = Asymmetry
k’ = Capacity factor
Figure 2Calibration curve for Doxophylline (for the UV-spectrophotometric method)
Figure 3Calibration curve for Doxophylline (high-performance liquid chromatography)
Characteristics of the Analytical method derived from the standard calibration curve (for UV-spectrophotometric method)
| Compound | LOD μg/ml | LOQ μg/ml | Linearity range μg/ml | Correlation coefficient r2 | Residual standard regression σ | Slope of regression S |
|---|---|---|---|---|---|---|
| Doxophylline | 0.07 | 0.2 | 4–30 | 0.99798 | 0.02569 | 0.00332 |
LOD = Limit of detection
LOQ = Limit of quantification
Characteristics of the analytical method derived from the standard calibration curve (for HPLC method)
| Compound | LOD μg/ml | LOQ μg/ml n = 5 | Linearity range μg/ml | Correlation coefficient r2 | Residual standard regression σ | Slope of regression S |
|---|---|---|---|---|---|---|
| Doxophylline | 0.05 | 0.165 | 4–30 | 0.99629 | 27483.89232 | 26167.08537 |
LOD = Limit of detection
LOQ = Limit of quantification
Figure 4Spectrum of Doxophylline (20 μg/ml) in 0.1 N hydrochloric acid by the ultravioled-visible spectrophotometer
Figure 5Chromatogram of the standard solution
Figure 6Chromatogram of the sample solution
Method precision (for UV-spectrophotometric method)
| Compound | Concentration μg/ml (n = 6) | Absorbance Mean ± SEM (n = 6) | % assay Mean ± SEM (n = 6) | % RSD (n = 6) |
|---|---|---|---|---|
| Doxophylline | 20 | 0.716 ± 0.00204 | 101.75 ± 0.3535 | 0.9 |
Method precision (for HPLC method)
| Compound | Concentration μg/ml (n = 6) | Retention time Mean ± SEM (n = 6) | % assay Mean ± SEM (n = 6) | % RSD (n = 6) |
|---|---|---|---|---|
| Doxophylline | 20 | 5.62 ± 0.0460 | 101.0 ± 0.4232 | 1.0 |
Method accuracy (for UV–spectrophotometric method)
| Level | Drug added (mg) | Drug recovered (mg) | % assay (Mean ± SEM) ( | % RSD of assay ( |
|---|---|---|---|---|
| Doxophylline | ||||
| 50% | 62.05 | 62.41 | 100.6 ± 0.088 | 0.2 |
| 100% | 124.01 | 123.75 | 99.8 ± 0.409 | 0.7 |
| 150% | 185.88 | 186.23 | 100.3 ± 0.266 | 0.5 |
Method accuracy (for HPLC method)
| Level | Drug added (mg) | Drug recovered (mg) | %assay (Mean ± SEM) (n = 3) | % RSD of assay (n = 3) |
|---|---|---|---|---|
| Doxophylline | ||||
| 50% | 62.05 | 62.21 | 100.26 ± 0.448 | 0.8 |
| 100% | 124.01 | 123.41 | 99.5 ± 0.458 | 0.8 |
| 150% | 185.88 | 186.27 | 100.2 ± 0.321 | 0.6 |
Method ruggedness (for UV-spectrophotometric method)
| Compound | % assay Mean ± SEM (n = 6) | % RSD of assay (n = 6) |
|---|---|---|
| Day 1, Analyst-1, Instrument-1 Doxophylline | 101.75 ± 0.3535 | 0.9 |
| Day 2, Analyst-2, Instrument-2 Doxophylline | 101.01 ± 0.1973 | 0.5 |
Method ruggedness (for HPLC method)
| Compound | % assay Mean ± SEM (n = 6) | % RSD of assay (n = 6) |
|---|---|---|
| Day 1, Analyst-1, Instrument-1, Column-1 Doxophylline | 101.0 ± 0.4232 | 1.0 |
| Day 2, Analyst-2, Instrument-2, Column-2 Doxophylline | 100.05 ± 0.20125 | 0.5 |
Method robustness (for HPLC method
| Compound | % RSD in normal | Changed condition (n = 5) | |
|---|---|---|---|
| Temperature | % RSD normal | % RSD (-5°C) | % RSD (+5°C) |
| Doxophylline | 1.0 | 0.18 | 1.05 |
| pH | % RSD normal | % RSD (-0.2 unit) | % RSD (+0.2 unit) |
| Doxophylline | 1.0 | 0.19 | 0.55 |
| Flow rate | % RSD normal | % RSD (-10%) | % RSD (+10%) |
| Doxophylline | 1.0 | 0.10 | 0.19 |
| Mobile phase ratio | % RSD normal | % RSD (-2%) | % RSD (+2%) |
| Doxophylline | 1.0 | 0.08 | 0.19 |
Comparison between UV and HPLC Method
| Parameter | UV method | HPLC method |
|---|---|---|
| Mechanism | Measurement of absorbance of samples containing only one absorbing component | Measurement of absorbance and separation (partition) of samples containing more than one absorbing component at a time |
| Accuracy and precision | Low compared to the RP-HPLC method | Very accurate and precise |
| Cost of analysis | Very low | High |
| Reagents/solvents/diluents/mobile phase | Use of a polar solvent generally is sufficient | Use of mobile phase having a combination of either buffer and polar solvent and/or use of two polar solvents |
| Analysis of compounds | Polar substances having λmax between 200 and 400 nm | Substances can be analyzed beyond the limit provided in the UV method due to the wider variety of the detector being employed |
| Sensitivity | Limited in sensitivity | Greater sensitivity (as various detectors can be employed) |
| Instrumentation | Easy to operate | Compared to the UV method, complex to operate |
| Speed | Compared to HPLC, analysis can be completed within lesser time | Time required for analysis depends on the nature of the molecule to be analyzed |
| Resolution | Low resolution compared to the HPLC method. Required to go for first and second derivative spectrophotometric methods | Greater resolution (wide variety of stationary phases) |
| Type of test/analysis | It can be use as a confirmatory test for a particular compound | It is used as a specific identification test for a particular compound |
| Useful at scale | Useful at laboratory scale at the primary level | Useful at a large scale, where complex molecules have to be analyzed |
| Applications | Useful to find out the qualitative parameter, like λmax of a particular compound | Useful to find out the quantitative parameters, like retention time of a particular compound |
| Degradation/by products | Can be analyzed simultaneously | Can be analyzed within one analysis |
| Calculation | Calculations have to be performed manually based on the λmax of a particular compound | Calculations are performed by the integrator itself |