The effects of height and surface roughness of abutments and the type of cement on bond strength of cement-retained implant restorations.

Khaled Q Al Hamad1, Bashar A Al Rashdan, Ethar H Abu-Sitta.   

Abstract

OBJECTIVES: To evaluate the effects and interactions between cement type, abutment height and surface roughness on bond strength of cemented implant restorations.
MATERIALS AND METHODS: One hundred and sixty metal copings were fabricated and divided into 16 groups of 10 samples each. Copings were cemented on 4 and 6 mm height Alfa-Gate(®) abutment, using four types of cements: zinc phosphate (ZNP), glass ionomer (GI), and zinc oxide eugenol (ZOE) with or without 15% vaseline. Copings were removed using a universal testing machine and bond strengths were recorded. All abutments were sandblasted with 50 μm aluminum oxide and the experiment was repeated. Results were analyzed using univariate analysis, Games-Howell, and Bonferroni's pairwise comparisons tests at P≤0.05.
RESULTS: Bond strengths were significantly different according to cement type, abutment height, and surface roughness (P=0.001). The cement ranking from highest to lowest was: ZNP>GI>ZOE>zinc oxide eugenol with 15% vaseline (ZOEV). An increasing abutment height showed a significant increase in bond strength for permanent cements only (P≤0.05). Sandblasting abutment surface significantly the increased bond strength for all cement except the ZOEV at 4 mm abutment, and for only ZOE and GI cements at the 6 mm abutment (P≤0.05).
CONCLUSIONS: ZP was the strongest cement and required using one of the variables (height, surface roughness) for maximum bond strength. GI bond strength was increased by both variables but sandblasting was more effective. ZOE required using both variables to be as effective as some of the permanent cement subgroups. ZOEV was the weakest and bond strength was not improved by either variables.
© 2010 John Wiley & Sons A/S.

Entities:  

Mesh:

Substances:

Year:  2010        PMID: 21039897     DOI: 10.1111/j.1600-0501.2010.02011.x

Source DB:  PubMed          Journal:  Clin Oral Implants Res        ISSN: 0905-7161            Impact factor:   5.977


  11 in total

1.  Comparative evaluation of the wear resistance of two different implant abutment materials after cyclic loading - An in vitro study.

Authors:  Maniamuthu Ragupathi; Vallabh Mahadevan; N S Azhagarasan; Hariharan Ramakrishnan; S Jayakrishnakumar
Journal:  Contemp Clin Dent       Date:  2020-11-26

2.  Evaluation of the Effect of Axial Wall Modification and Coping Design on the Retention of Cement-retained Implant-supported Crowns.

Authors:  Reza Derafshi; Ahmad Hasan Ahangari; Kianoosh Torabi; Mitra Farzin
Journal:  J Dent Res Dent Clin Dent Prospects       Date:  2015-03-04

3.  Effect of abutment modification and cement type on retention of cement-retained implant supported crowns.

Authors:  Mitra Farzin; Kianoosh Torabi; Ahmad Hasan Ahangari; Reza Derafshi
Journal:  J Dent (Tehran)       Date:  2014-05-31

4.  Effect of Surface Modifications on the Retention of Cement-retained Implant Crowns under Fatigue Loads: An In vitro Study.

Authors:  R Ajay; K Suma; Seyed Asharaf Ali; Jambai Sampath Kumar Sivakumar; V Rakshagan; V Devaki; K Divya
Journal:  J Pharm Bioallied Sci       Date:  2017-11

5.  Zinc phosphate as a definitive cement for implant-supported crowns and fixed dentures.

Authors:  Dennis Flanagan
Journal:  Clin Cosmet Investig Dent       Date:  2017-11-03

6.  Effect of Implant Abutment Acid Etching on the Retention of Crowns Luted with Different Cements: An In Vitro Comparative Evaluation.

Authors:  R Ajay; V Rakshagan; M Kamatchi; A SelvaBalaji; Jambai Sampath Kumar Sivakumar; M Senthil Kumar
Journal:  J Pharm Bioallied Sci       Date:  2019-05

7.  An in vitro study to compare the influence of newer luting cements on retention of cement-retained implant-supported prosthesis.

Authors:  Hasan Sarfaraz; Arifa Hassan; K Kamalakanth Shenoy; Mallika Shetty
Journal:  J Indian Prosthodont Soc       Date:  2019 Apr-Jun

8.  Marginal Discrepancy of Cast Copings to Abutments with Three Different Luting Agents.

Authors:  José Rosas; Frank Mayta-Tovalino; Maria Eugenia Guerrero; Pedro Luis Tinedo-López; Claudia Delgado; Vanessa Z Ccahuana-Vasquez
Journal:  Int J Dent       Date:  2019-09-02

9.  Effect of surface treatments on the retention of implant-supported cement-retained bridge with short abutments: An in vitro comparative evaluation.

Authors:  Monica Shrivastav
Journal:  J Indian Prosthodont Soc       Date:  2018 Apr-Jun

10.  In Vitro Impact Testing to Simulate Implant-Supported Prosthesis Retrievability in Clinical Practice: Influence of Cement and Abutment Geometry.

Authors:  Andrea T Lugas; Mara Terzini; Elisabetta M Zanetti; Gianmario Schierano; Carlo Manzella; Domenico Baldi; Cristina Bignardi; Alberto L Audenino
Journal:  Materials (Basel)       Date:  2020-04-09       Impact factor: 3.623

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.