Literature DB >> 21034096

Stabilization of sample temperature in a surface-science vacuum chamber to 0.03 K and quartz-crystal microbalance frequency to 0.06 Hz over 0.5 h.

Alan J Slavin1.   

Abstract

Improvements have been made to a high-stability quartz-crystal microbalance for use in a typical surface-science, ultrahigh vacuum chamber, with a frequency stability of one part in 10(8) (0.06 Hz) over 0.5 h. This gives a resolution equivalent to 2% of an atomic monolayer of oxygen over 0.5 h. The quartz-crystal microbalance (QCM) crystal can be rotated to different surface-analysis positions in the chamber. These characteristics open up the combination of surface and bulk adsorption studies on the same sample without transferring the sample to another chamber. To accomplish this, it was necessary to stabilize the sample temperature to ±0.03 K over several hours. The oscillator performance is illustrated by the uptake of oxygen by a gold-plated QCM crystal.

Entities:  

Year:  2010        PMID: 21034096     DOI: 10.1063/1.3488368

Source DB:  PubMed          Journal:  Rev Sci Instrum        ISSN: 0034-6748            Impact factor:   1.523


  1 in total

1.  One-Port Electronic Detection Strategies for Improving Sensitivity in Piezoelectric Resonant Sensor Measurements.

Authors:  Zhongxu Hu; John Hedley; Neil Keegan; Julia Spoors; Barry Gallacher; Calum McNeil
Journal:  Sensors (Basel)       Date:  2016-10-25       Impact factor: 3.576

  1 in total