Literature DB >> 21034092

Quantitative scanning probe microscope topographies by charge linearization of the vertical actuator.

Andrew J Fleming1.   

Abstract

Many forms of scanning probe microscopy require a piezoelectric actuator to vary the probe-sample distance. Examples include constant-force atomic force microscopy and constant-current scanning tunneling microscopy. In such modes, the topography of the sample is reconstructed from the voltage applied to the vertical piezoelectric actuator. However, piezoelectric actuators exhibit significant hysteresis which can produce up to 14% uncertainty in the reproduced topography. In this work, a charge drive is used to linearize the vertical piezoelectric actuator which reduces the error from 14% to 0.65%.

Mesh:

Year:  2010        PMID: 21034092     DOI: 10.1063/1.3488359

Source DB:  PubMed          Journal:  Rev Sci Instrum        ISSN: 0034-6748            Impact factor:   1.523


  1 in total

1.  Voice coil-based scanning probe microscopy.

Authors:  Petr Klapetek; Miroslav Valtr; Václav Ducho 328; Jaroslav Sobota
Journal:  Nanoscale Res Lett       Date:  2012-06-21       Impact factor: 4.703

  1 in total