A fast and simple assay for busulfan in serum or plasma by liquid chromatography-tandem mass spectrometry using turbulent flow online extraction technology.

Dustin R Bunch1, Courtney Heideloff, James C Ritchie, Sihe Wang.   

Abstract

Busulfan is used in myeloablative preparation regimens for hematopoietic bone marrow transplantation. Due to its narrow therapeutic range therapeutic drug monitoring of busulfan is recommended. In this study a fast and simple method for measuring busulfan in serum or plasma by liquid chromatography-tandem mass spectrometry (LC-MS/MS) has been developed utilizing turbulent flow online extraction technology. Serum or plasma was mixed with acetonitrile containing d(8)-busulfan. After centrifugation the supernatant was injected onto a turbulent flow preparatory column then transferred to a C18 analytical column monitored by a tandem mass spectrometer set at positive electrospray ionization. The analytical cycle time was 4.0min. The method was linear from 0.15 to 41.90μmol/L with an accuracy of 87.9-103.0%. Inter- and intra-assay CVs across four concentration levels were 2.1-7.8%. No significant carryover or ion suppression was observed. No interference was observed from commercial control materials containing more than 100 compounds. Comparison with a well established LC-MS/MS method using patient specimens (n=45) showed a mean bias 1.3% with Deming regression of slope 1.02, intercept -0.02μmol/L, and a linear correlation coefficient 0.9883. The LC-MS/MS method coupled with turbulent flow online sample cleaning technology described here offers reliable busulfan quantitation in serum or plasma with minimum manual sample preparation and was fully validated for clinical use.
Copyright © 2010 Elsevier B.V. All rights reserved.

Entities:  

Mesh:

Substances:

Year:  2010        PMID: 21030324     DOI: 10.1016/j.jchromb.2010.10.003

Source DB:  PubMed          Journal:  J Chromatogr B Analyt Technol Biomed Life Sci        ISSN: 1570-0232            Impact factor:   3.205


  3 in total

1.  An LC-MS Assay with Isocratic Separation and On-line Solid Phase Extraction to Improve the Routine Therapeutic Drug Monitoring of Busulfan in Plasma.

Authors:  Cristiano Ialongo; Alessia Francesca Mozzi; Sergio Bernardini
Journal:  J Med Biochem       Date:  2017-04-22       Impact factor: 3.402

2.  The successful implementation of a licensed data management interface between a Sunquest(®) laboratory information system and an AB SCIEX™ mass spectrometer.

Authors:  Deborah French; Enrique Terrazas
Journal:  J Pathol Inform       Date:  2013-01-31

3.  Quantification of human plasma-busulfan concentration by liquid chromatography-tandem mass spectrometry.

Authors:  Soo Young Moon; Min Kyoo Lim; Susie Hong; Yongbum Jeon; Minje Han; Sang Hoon Song; Kyoung Soo Lim; Kyung-Sang Yu; In-Jin Jang; Ji Won Lee; Hyoung Jin Kang; Junghan Song
Journal:  Ann Lab Med       Date:  2013-12-06       Impact factor: 3.464

  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.