Literature DB >> 21030205

Biceps tenodesis with interference screw fixation: a biomechanical comparison of screw length and diameter.

Mark A Slabaugh1, Rachel M Frank, Geoffrey S Van Thiel, Rebecca M Bell, Vincent M Wang, Scott Trenhaile, Matthew T Provencher, Anthony A Romeo, Nikhil N Verma.   

Abstract

PURPOSE: To evaluate the effect of screw length and diameter on the mechanical properties of biceps tenodesis (BT) with an interference screw in 2 different locations (proximal and distal).
METHODS: We randomized 42 fresh-frozen human cadaveric shoulders (mean age, 65 ± 8 years) into 6 groups (n = 7): arthroscopic proximal BT using 7 × 15-, 7 × 25-, 8 × 15-, or 8 × 25-mm interference screws or distal subpectoral BT with 7 × 15- or 8 × 15-mm interference screws. Each repaired specimen was mounted onto a materials testing machine, preloaded to 5 N for 2 minutes, cycled from 5 to 70 N for 500 cycles (1 Hz), and loaded to failure (1 mm/s). Displacement during cyclical loading, pullout stiffness, and ultimate load to failure were computed, and the mechanism of failure was noted.
RESULTS: All failures occurred at the tendon-screw interface. There was no statistically significant difference in ultimate displacement among all groups in the ultimate load to failure, displacement at peak load, and stiffness.
CONCLUSIONS: There is no difference in ultimate load to failure, displacement at peak load, and stiffness of BT with regard to screw length or diameter at both proximal and distal tenodesis locations. These data would support use of a smaller-diameter and shorter implant for BT both proximally and distally. CLINICAL RELEVANCE: The results may serve as a guide to the orthopaedic surgeon performing proximal BT in selecting the appropriate interference screw. When possible, we recommend using the smallest screw size available to minimize risk of stress fracture at the tenodesis site. Copyright Â
© 2011 Arthroscopy Association of North America. Published by Elsevier Inc. All rights reserved.

Entities:  

Mesh:

Year:  2010        PMID: 21030205     DOI: 10.1016/j.arthro.2010.07.004

Source DB:  PubMed          Journal:  Arthroscopy        ISSN: 0749-8063            Impact factor:   4.772


  10 in total

1.  Biomechanical characterization of unicortical button fixation: a novel technique for proximal subpectoral biceps tenodesis.

Authors:  Joseph P DeAngelis; Alvin Chen; Michael Wexler; Benjamin Hertz; Leandro Grimaldi Bournissaint; Ara Nazarian; Arun J Ramappa
Journal:  Knee Surg Sports Traumatol Arthrosc       Date:  2013-11-20       Impact factor: 4.342

2.  Arthroscopic Biceps Tenodesis From a Superior Viewing Portal in the Shoulder.

Authors:  Andrew A Tarleton; Liang Zhou; Michael J O'Brien; Felix H Savoie
Journal:  Arthrosc Tech       Date:  2015-08-24

3.  The biomechanical performance of a new forked knotless biceps tenodesis compared to a knotless and suture anchor tenodesis.

Authors:  Olaf Lorbach; Christian Trennheuser; Dieter Kohn; Konstantinos Anagnostakos
Journal:  Knee Surg Sports Traumatol Arthrosc       Date:  2014-10-08       Impact factor: 4.342

4.  "Relaxed" biceps proximal tenodesis: an arthroscopic technique with decreased residual tendon tension.

Authors:  Philippe Valenti; Ivan Benedetto; Ali Maqdes; Sara Lima; Constantina Moraiti
Journal:  Arthrosc Tech       Date:  2014-10-27

Review 5.  The rate and reporting of fracture after biceps tenodesis: A systematic review.

Authors:  Hailey P Huddleston; Joey S Kurtzman; Samuel Gedailovich; Steven M Koehler; William R Aibinder
Journal:  J Orthop       Date:  2021-11-24

Review 6.  Management of proximal biceps tendon pathology.

Authors:  Simon P Lalehzarian; Avinesh Agarwalla; Joseph N Liu
Journal:  World J Orthop       Date:  2022-01-18

7.  Biomechanical Comparison of a Novel Multiplanar, Perpendicular Whipstitch With the Krackow Stitch and Standard Commercial Whipstitch.

Authors:  Stefano Muscatelli; Kempland C Walley; Conor S Daly-Seiler; Joseph A Greenstein; Aaron Sciascia; David P Patterson; Michael T Freehill
Journal:  Orthop J Sports Med       Date:  2022-08-12

8.  Clinical and Biomechanical Evaluation of an All-Arthroscopic Suprapectoral Biceps Tenodesis.

Authors:  Cynthia A Kahlenberg; Ronak M Patel; Rueben Nair; Prashant P Deshmane; Galen Harnden; Michael A Terry
Journal:  Orthop J Sports Med       Date:  2014-10-23

9.  Biomechanical evaluation of the long head of the biceps brachii tendon fixed by three techniques: a sheep model.

Authors:  Carlos Henrique Ramos; Júlio Cezar Uili Coelho
Journal:  Rev Bras Ortop       Date:  2016-12-30

10.  Modified Open Suprapectoral EndoButton Tension Slide Tenodesis Technique of Long Head of Biceps with Restored Tendon Tension-Length.

Authors:  Jagadish Prabhu; Mohammed Khalid Faqi; Rashad Khamis Awad; Fahad Alkhalifa
Journal:  Open Orthop J       Date:  2017-03-31
  10 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.