Comparative analysis of white-to-white and angle-to-angle distance measurements with partial coherence interferometry and optical coherence tomography.

Gabor Nemeth1, Ziad Hassan, Eszter Szalai, Andras Berta, Laszlo Modis.   

Abstract

PURPOSE: To perform white-to-white (WTW) and angle-to-angle (ATA) distance measurements with 2 optical devices and analyze the correlation with other anterior segment parameters.
SETTING: Department of Ophthalmology, Medical Health and Science Center, University of Debrecen, Debrecen, Hungary.
DESIGN: Comparative case series.
METHODS: In this prospective study, the WTW distance, axial length (AL), and anterior chamber depth (ACD) were measured by partial coherence interferometry (PCI) (IOLMaster). The horizontal and vertical ATA distances and ACD were assessed by anterior segment optical coherence tomography (AS-OCT) (Visante).
RESULTS: The mean PCI values were AL, 23.31 mm ± 1.52 (SD); WTW distance, 11.99 ± 0.47 mm; ACD, 2.99 ± 0.4 mm. On AS-OCT, the mean horizontal ATA was 11.43 ± 0.51 mm and the mean vertical ATA, 10.72 ± 0.66 mm. The WTW distance was significantly greater than the horizontal ATA distance, and the horizontal ATA distance was significantly greater than the vertical ATA distance (P<.0001). The correlations between WTW and horizontal ATA (r = 0.51; P<.0001), vertical ATA (r = 0.32; P<.0021), AL (r = 0.45; P<.0001), and patient age (r = -0.28; P = .023) were significant. The PCI WTW distance and ACD were significantly correlated (r = 0.55; P<.0001). There were no significant correlations between AS-OCT horizontal and vertical ATA measurements (r = -0.02; P = 0.79) or between same-device ACD measurements (r = 0.18; P = .15).
CONCLUSIONS: Although PCI WTW and AS-OCT ATA measurements are not clinically interchangeable, the WTW values correlated well with horizontal and vertical measurements and other anterior segment parameters.
Copyright © 2010 ASCRS and ESCRS. Published by Elsevier Inc. All rights reserved.

Entities:  

Mesh:

Year:  2010        PMID: 21029893     DOI: 10.1016/j.jcrs.2010.05.017

Source DB:  PubMed          Journal:  J Cataract Refract Surg        ISSN: 0886-3350            Impact factor:   3.351


  8 in total

1.  Evaluation of anterior chamber parameters with spectral-domain optical coherence tomography.

Authors:  Isabel Pinilla Lozano; Carmen López de la Fuente; Francisco Segura; Elvira Orduna Hospital; Ana Sánchez-Cano
Journal:  Jpn J Ophthalmol       Date:  2017-11-29       Impact factor: 2.447

2.  Angle-to-angle and spur-to-spur distance analysis with high-resolution optical coherence tomography.

Authors:  Robert Montés-Micó; Pedro Tañá-Rivero; Salvador Aguilar-Córcoles; María Ruiz-Santos; María Dolores Rodríguez-Carrillo; Ramón Ruiz-Mesa
Journal:  Eye Vis (Lond)       Date:  2020-08-16

3.  White-to-white corneal diameter distribution in an adult population.

Authors:  Hassan Hashemi; Mehdi Khabazkhoob; Mohammad Hassan Emamian; Mohammad Shariati; Abbasali Yekta; Akbar Fotouhi
Journal:  J Curr Ophthalmol       Date:  2015-10-19

4.  Estimation of axial curvature of anterior sclera: correlation between axial length and anterior scleral curvature as affected by angle kappa.

Authors:  Sang-Mok Lee; Hyuk Jin Choi; Heejin Choi; Mee Kum Kim; Won Ryang Wee
Journal:  BMC Ophthalmol       Date:  2016-10-07       Impact factor: 2.209

5.  Three-dimensional non-parametric method for limbus detection.

Authors:  Ahmed Abass; Bernardo T Lopes; Ashkan Eliasy; Richard Wu; Steve Jones; John Clamp; Renato Ambrósio; Ahmed Elsheikh
Journal:  PLoS One       Date:  2018-11-26       Impact factor: 3.240

6.  Correlation between intraoperative and postoperative vaulting of the EVO implantable Collamer lens: a retrospective study of real-time observations of vaulting using the RESCAN 700 system.

Authors:  Nian Guan; Xiao-Nong Zhang; Wan-Jun Zhang
Journal:  BMC Ophthalmol       Date:  2022-01-03       Impact factor: 2.209

7.  Comparative analysis of two optical biometry devices: high wavelength swept source OCT versus partial coherence interferometry.

Authors:  Eszter Szalai; Adrienne Csutak
Journal:  Int Ophthalmol       Date:  2021-10-11       Impact factor: 2.029

8.  Daily Variations of Corneal White-to-White Diameter Measured with Different Methods.

Authors:  Stefan Bandlitz; Makram Nakhoul; Konstantin Kotliar
Journal:  Clin Optom (Auckl)       Date:  2022-09-20
  8 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.