Literature DB >> 2100174

Techniques of intraocular pressure determination.

K Green1.   

Abstract

A variety of indentation or applanation tonometers are available for the measurement of intraocular pressure in conscious or anesthetized animals. The Schiøtz tonometer is a simple indentation tonometer, but unless used properly can give misleading values. Applanation tonometers include the air tonometer, pneumatonograph, and the Goldmann type (either slit-lamp or hand-held models). These flatten a small, fixed area of cornea and measure the resultant pressure. In anesthetized animals anterior chamber cannulation may be performed, although breakdown of the blood-aqueous barrier can ensue together with an alteration of intraocular pressure. Of importance is that all tonometers be calibrated for each species since scleral rigidity, ocular curvature, pressure/volume relationships, etc. all vary between species.

Entities:  

Mesh:

Year:  1990        PMID: 2100174

Source DB:  PubMed          Journal:  Lens Eye Toxic Res        ISSN: 1042-6922


  1 in total

1.  Underestimation of intraocular pressure after photorefractive keratectomy: a biomechanical analysis.

Authors:  Per Hallberg; Anders Eklund; Kenneth Santala; Timo Koskela; Olof Lindahl; Christina Lindén
Journal:  Med Biol Eng Comput       Date:  2006-07-26       Impact factor: 2.602

  1 in total