Literature DB >> 20977825

Large-diameter graft-stent (Advanta V12) implantation in various locations: early results.

Dietmar Schranz1, Christian Jux, Melanie Vogel, Jürgen Bauer, Hakan Akintürk, Klaus Valeske.   

Abstract

OBJECTIVES: Transcatheter stent placement carries the risk of cardiovascular aneurysm or rupture. Covered stent implantation reduces these risks. The recently marketed Advanta V12 large-diameter-covered stent is pre-mounted and requires 9 (8)-11 Fr delivery systems. The aim was to report on the early results of the treatment of various cardiovascular obstructions by the implantation of a new polytetrafluoroethylene-covered stent (V12).
METHODS: Graft stents on balloons with a diameter (12, 14, 16 millimetres) sufficient to anchor the stent in various obstructions (congenital aortic coarctation, n = 5; obstruction after ascending aorta repair, n = 2; pulmonary arteries, n = 5; inferior caval vein, n = 1; atretic superior caval vein, n = 1; pulmonary vein obstruction, n = 1; and right ventricular outflow tract, n = 1) were implanted using the smallest available delivery system. Secondary dilation with larger-diameter balloons was performed when the residual pressure was gradient, the stent-vessel wall relationship or stent re-coiling due to different reasons needed a re-intervention by pure ballooning or second stent placement.
RESULTS: All 16 patients aged 5-46 years underwent V12 implantation. The variability of the treated lesions and the need for additional interventions were responsible for large ranges in fluoroscopy time between 7.3 to 48.2 minutes (median 17.3). Considering the additional procedures, the V12 stent achieved the desired result in all cases. There were no major complications. At short-term median follow-up of 2 months, all patients are alive and well with no evidence of stent failing.
CONCLUSION: These initial results show that the covered Advanta V12 large-diameter stent is safe and effective in the immediate treatment of various cardiovascular obstructions. Long-term follow-up is required.

Entities:  

Mesh:

Year:  2010        PMID: 20977825     DOI: 10.1017/S1047951110001459

Source DB:  PubMed          Journal:  Cardiol Young        ISSN: 1047-9511            Impact factor:   1.093


  4 in total

Review 1.  Covered Stents in the Management of Aortic Coarctation and Right Ventricular Outflow Tract Obstruction.

Authors:  Nibras E El Sherif; Nathaniel W Taggart
Journal:  Curr Cardiol Rep       Date:  2022-01-13       Impact factor: 2.931

2.  The use of expanded polytetrafluoroethylene in depressed deformities of the face.

Authors:  Xiaowei Liu; Yiming Zhang; Shaoliang Wang; Zeyuan Lei; Xiang Li; Dongli Fan
Journal:  Exp Ther Med       Date:  2016-10-03       Impact factor: 2.447

3.  Iliac occlusion due to covered stent deformation following abdominal massages.

Authors:  Gaia Pollorsi; Daniel Danzer; Nicolas Murith; Christoph Huber; Damiano Mugnai
Journal:  J Vasc Surg Cases Innov Tech       Date:  2022-03-09

4.  Middle Aortic Syndrome Treated by Implantation of an Advanta V12 Large Diameter Stent.

Authors:  Meng-Luen Lee; Ing-Sh Chiu; Albert D Yang
Journal:  Arq Bras Cardiol       Date:  2016-01       Impact factor: 2.000

  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.