Literature DB >> 20971374

Planning implants crown down--a systematic quality control for proof of concept.

Rolf Ewers1, Rudolf Seemann, Gerald Krennmair, Kurt Schicho, Ayman Omar Kurdi, Axel Kirsch, Astrid Reichwein.   

Abstract

PURPOSE: We evaluated an innovative approach for conventional surgical or transmucosal implantation applying backward planning with bone oriented crown down implant positions, followed by immediate restoration.
MATERIALS AND METHODS: Our workflow combined computer-assisted preoperative planning with the well-known intraoperative handling of surgical templates. To guarantee optimal accuracy, the complete process was bone borne, not soft tissue borne. Furthermore, in this concept, the definite prosthesis was manufactured on a plaster cast and fixed with technical implants inserted into the model using the same drilling template applied for intraoperative guidance of the drill. This bone-based workflow avoids the accumulation of errors throughout all steps of the treatment. In the present study, we have provided a concise quantitative quality control using 4 in vivo implants. This goal was accomplished by image matching of the preoperative plan with the corresponding postoperative computed tomography scan, applying descriptive statistics, the 1-sample t test, and the Student t test.
RESULTS: Statistical evaluation showed that the bone-borne drilling template provides sufficient accuracy for immediate restoration with the definite prosthesis at the end of the operation. Deviation in the apex plane was 0.8 ± 0.42 mm, with 0.33 at the maximum. The angular deviation was 3.18° ± 0.78°.
CONCLUSIONS: Our workflow is a promising approach with respect to handling, accuracy, and intraoperative safety.
Copyright © 2010 American Association of Oral and Maxillofacial Surgeons. Published by Elsevier Inc. All rights reserved.

Entities:  

Mesh:

Substances:

Year:  2010        PMID: 20971374     DOI: 10.1016/j.joms.2009.03.024

Source DB:  PubMed          Journal:  J Oral Maxillofac Surg        ISSN: 0278-2391            Impact factor:   1.895


  1 in total

1.  Concept description and accuracy evaluation of a moldable surgical targeting system.

Authors:  Thomas S Rau; Sina Witte; Lea Uhlenbusch; Lüder A Kahrs; Thomas Lenarz; Omid Majdani
Journal:  J Med Imaging (Bellingham)       Date:  2021-02-19