Literature DB >> 20964208

Overcorrection of iodinated contrast attenuation in SPECT-CT: phantom studies.

Dacian V Bonta1, Richard L Wahl.   

Abstract

PURPOSE: The authors investigated the effects of iodinated contrast agents on the quantification of radioactivity obtained using a combined single photon emission tomography (SPECT) and high resolution 16 slice computed tomography (CT) scanner.
METHODS: Two conditions were evaluated: Contrast media mixed with radioisotope and the contrast media, and the radioisotope solution confined in adjacent but distinct volumes. Phantoms containing combinations of 99mTc, 111In, and 131I solutions in normal saline and contrast solutions in normal saline corresponding to the two conditions were prepared and scanned. Images were reconstructed by iterative reconstruction (ordered subset expectation maximization), with and without CT-derived attenuation maps. The reconstructed counts encoded in the reconstructed images were compared.
RESULTS: Compared to normal saline, the presence of an iodinated contrast agent resulted in the underestimation of reconstructed counts in all nonattenuation corrected SPECT images. Compared to normal saline, the presence of a contrast agent resulted in the overestimation of reconstructed counts in all attenuation corrected SPECT images. The largest underestimation of reconstructed counts was 12.8% and the largest overestimation value was 35.9%.
CONCLUSIONS: High concentrations of a contrast agent caused errors in the measurement of actual radiotracer reconstructed counts in phantoms. These errors could have important consequences for the use of oral and intravenous contrast in both diagnostic (qualitative) and dosimetric (quantitative) SPECT-CT imaging. The authors' findings suggest that noncontrast imaging or alternative contrast-specific attenuation correction approaches should be considered for optimal SPECT activity quantification.

Entities:  

Mesh:

Substances:

Year:  2010        PMID: 20964208     DOI: 10.1118/1.3483100

Source DB:  PubMed          Journal:  Med Phys        ISSN: 0094-2405            Impact factor:   4.071


  2 in total

1.  Intravenous contrast-enhanced CT can be used for CT-based attenuation correction in clinical (111)In-octreotide SPECT/CT.

Authors:  Thomas Levin Klausen; Jann Mortensen; Robin de Nijs; Flemming Littrup Andersen; Liselotte Højgaard; Thomas Beyer; Søren Holm
Journal:  EJNMMI Phys       Date:  2015-02-12

Review 2.  Best practice for the nuclear medicine technologist in CT-based attenuation correction and calcium score for nuclear cardiology.

Authors:  Luca Camoni; Andrea Santos; Marieclaire Attard; Marius Ovidiu Mada; Agata Karolina Pietrzak; Sonja Rac; Sebastijan Rep; Christelle Terwinghe; Pedro Fragoso Costa
Journal:  Eur J Hybrid Imaging       Date:  2020-07-06
  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.