Literature DB >> 20962192

Metal-on-metal hip arthroplasty in patients thirty years of age or younger.

Julien Girard1, Donatien Bocquet, Guillaume Autissier, Nicolas Fouilleron, Damien Fron, Henri Migaud.   

Abstract

BACKGROUND: Total hip arthroplasty in patients younger than thirty years of age represents a long-term challenge. As polyethylene wear secondary to a high activity level could be problematic, hard-on-hard bearings have been proposed to reduce wear. The aim of this retrospective case series was to assess the clinical and radiographic results of primary metal-on-metal total hip arthroplasty in patients thirty years of age or younger.
METHODS: We retrospectively studied thirty-four patients (forty-seven hips) who had undergone metal-on-metal total hip arthroplasty and analyzed the radiographic and clinical measurements after a mean duration of follow-up of 108 months (range, 62.4 to 153.6 months). The mean age of the patients at the time of surgery was twenty-five years (range, fifteen to thirty years). The diameter of the head of the femoral component was 28 mm in all hips except five, in which it was 32 mm. The metal-on-metal bearing was the same in all patients.
RESULTS: The mean Merle d'Aubigné score increased from 10.6 (range, 1 to 14) to 17.1 (range, 12 to 18). No wear was found on the latest radiograph, but osteolysis was noted in three femora and two acetabula. Two revisions were performed, one because of impingement secondary to cup malorientation and the other because of acetabular loosening with osteolysis. Kaplan-Meier analysis with revision of either component as the end point revealed a ten-year survival rate of 94.5% (95% confidence interval, 80% to 98.6%). The survival rate of the femoral stem was 100%.
CONCLUSIONS: These encouraging intermediate-term results indicate that hip arthroplasty with metal-on-metal bearing components may be a suitable solution for young and active patients with hip osteoarthritis or osteonecrosis of the femoral head.

Entities:  

Mesh:

Substances:

Year:  2010        PMID: 20962192     DOI: 10.2106/JBJS.I.01644

Source DB:  PubMed          Journal:  J Bone Joint Surg Am        ISSN: 0021-9355            Impact factor:   5.284


  25 in total

1.  Correlation between groin pain and cup design of hip-resurfacing implants: a prospective study.

Authors:  Julien Girard; Erwan Pansard; Reda Ouahes; Henri Migaud; Cyril Delay; Laurent Vasseur
Journal:  Int Orthop       Date:  2013-12-10       Impact factor: 3.075

2.  Stability and trunnion wear potential in large-diameter metal-on-metal total hips: a finite element analysis.

Authors:  Jacob M Elkins; John J Callaghan; Thomas D Brown
Journal:  Clin Orthop Relat Res       Date:  2014-02       Impact factor: 4.176

3.  Edge-loading severity as a function of cup lip radius in metal-on-metal total hips--a finite element analysis.

Authors:  Jacob M Elkins; Karen M Kruger; Douglas R Pedersen; John J Callaghan; Thomas D Brown
Journal:  J Orthop Res       Date:  2011-08-02       Impact factor: 3.494

4.  Survivorship of hip and knee implants in pediatric and young adult populations: analysis of registry and published data.

Authors:  Art Sedrakyan; Lucas Romero; Stephen Graves; David Davidson; Richard de Steiger; Peter Lewis; Michael Solomon; Robyn Vial; Michelle Lorimer
Journal:  J Bone Joint Surg Am       Date:  2014-12-17       Impact factor: 5.284

Review 5.  Has total hip arthroplasty in patients 30 years or younger improved? A systematic review.

Authors:  Muyibat A Adelani; James A Keeney; Allison Palisch; Susan A Fowler; John C Clohisy
Journal:  Clin Orthop Relat Res       Date:  2013-04-06       Impact factor: 4.176

6.  High prevalence of adverse reactions to metal debris in small-headed ASR™ hips.

Authors:  Aleksi Reito; Timo Puolakka; Petra Elo; Jorma Pajamäki; Antti Eskelinen
Journal:  Clin Orthop Relat Res       Date:  2013-04-30       Impact factor: 4.176

Review 7.  What are the advantages and disadvantages of imaging modalities to diagnose wear-related corrosion problems?

Authors:  Denis Nam; Robert L Barrack; Hollis G Potter
Journal:  Clin Orthop Relat Res       Date:  2014-03-25       Impact factor: 4.176

8.  Alumina-on-alumina THA performed in patients younger than 30 years: a 10-year minimum followup study.

Authors:  Hyeong Jo Yoon; Jeong Joon Yoo; Kang Sup Yoon; Kyung-Hoi Koo; Hee Joong Kim
Journal:  Clin Orthop Relat Res       Date:  2012-07-24       Impact factor: 4.176

9.  Ceramic bearings for total hip arthroplasty have high survivorship at 10 years.

Authors:  James A D'Antonio; William N Capello; Marybeth Naughton
Journal:  Clin Orthop Relat Res       Date:  2012-02       Impact factor: 4.176

10.  Can patients return to high-impact physical activities after hip resurfacing? A prospective study.

Authors:  Julien Girard; Bruno Miletic; Anthony Deny; Henri Migaud; Nicolas Fouilleron
Journal:  Int Orthop       Date:  2013-03-02       Impact factor: 3.075

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.