| Literature DB >> 2095752 |
J H Wilson1, R Rath, R Glaser, T Panke.
Abstract
We report 2 cases of severe intravascular hemolysis after mitral valve repair using a Duran annuloplasty ring. In both patients residual mitral regurgitation was present, hemolysis was severe enough to warrant a second operation, and hemolysis ceased immediately after the second operation. We believe that a high-velocity regurgitant jet directed toward the cloth-covered annuloplasty ring was responsible for the hemolysis in both patients.Entities:
Mesh:
Year: 1990 PMID: 2095752 DOI: 10.1016/0003-4975(90)90108-i