Inaccuracy of radiological and ECG criteria for right ventricular outflow tract implantation of pacemaker leads.

Andrei-Dumitru Margulescu1, Roxana Cristina Sisu, Raluca Dulgheru, Dragos Vinereanu.   

Abstract

We present three cases of active fixation pacemaker lead implantation, where the leads were believed to be inserted in the right ventricular outflow tract (RVOT) septal position, using established radiological criteria. However, when the exact location of the leads was documented by three-dimensional echocardiography, true septal RVOT position was achieved in only one patient. In the other two patients, the leads were, in fact, implanted into the RVOT anterior free wall. The electrocardiogram criteria were not able to identify the septal position from the anterior RVOT free wall position.

Entities:  

Mesh:

Year:  2010        PMID: 20952766     DOI: 10.1093/europace/euq371

Source DB:  PubMed          Journal:  Europace        ISSN: 1099-5129            Impact factor:   5.214


  2 in total

1.  Inadequacy of fluoroscopy and electrocardiogram in predicting septal position in RVOT pacing - Validation with cardiac computed tomography.

Authors:  Gautam Sharma; Salman Salahuddin; Prashanthan Sanders; Himanshu Gupta; Gurpreet Gulati; Priya Jagia; V K Bahl
Journal:  Indian Heart J       Date:  2015-11-30

2.  Left ventricular strain analysis reveals better synchrony and diastolic function for septal versus apical right ventricular permanent pacing.

Authors:  Roxana Cristina Rimbas; Andrei Dumitru Margulescu; Calin Siliste; Dragos Vinereanu
Journal:  Maedica (Buchar)       Date:  2014-09