Literature DB >> 20952032

Joint AUA/SUNA white paper on reprocessing of flexible cystoscopes.

J Quentin Clemens1, Robert Dowling, Frances Foley, Howard B Goldman, Chris M Gonzalez, Christopher Tessier, Mary Anne Wasner, Eileen Young.   

Abstract

PURPOSE: Numerous federal bodies and professional societies have produced guidelines and standards for the reprocessing of endoscopes, but few specifically address the reprocessing of cystoscopes.
MATERIALS AND METHODS: This document presents a summary of the current recommendations for reprocessing flexible cystoscopes, and highlights particular aspects of instrument reprocessing that are unique to cystoscopy.
RESULTS: Cystoscopes are classified as "semi-critical" devices. Such devices require high-level disinfection or sterilization between patients. High-level disinfection differs from sterilization in that high-level disinfection does not kill large numbers of bacterial spores, while sterilization involves the complete destruction of all microbial life. In the office setting, high-level disinfection (using glutaraldehyde or another chemical disinfectant) is commonly employed for the reprocessing of flexible cystoscopes. During high-level disinfection, standard reprocessing steps should be followed to prevent contamination of the cystoscope between uses. These include precleaning, leak testing, cleaning, disinfection, rinsing, and drying. Recommended "soak times" for glutaraldehyde vary from 20 to 45 minutes, depending upon the degree to which these steps are followed. If no precleaning is performed, a 45-minute glutaraldehyde soak is required to achieve high-level disinfection. Conversely, a 20-minute soak is adequate to achieve high-level disinfection if recommended reprocessing steps are followed prior to immersion in the glutaraldehyde. One chemical disinfectant (ortho-phthalaldehyde) has been associated with anaphylaxis in bladder cancer patients, and should be avoided in these patients.
CONCLUSIONS: This white paper provides a concise reference document for the reprocessing of flexible cystoscopes. In addition, references and links to more comprehensive resources are provided. This document may be useful for clinicians and others who are in search of guidance in this area.
Copyright © 2010 American Urological Association Education and Research, Inc. Published by Elsevier Inc. All rights reserved.

Entities:  

Mesh:

Year:  2010        PMID: 20952032     DOI: 10.1016/j.juro.2010.08.001

Source DB:  PubMed          Journal:  J Urol        ISSN: 0022-5347            Impact factor:   7.450


  2 in total

1.  LithoVue™: A new single-use digital flexible ureteroscope.

Authors:  Salvatore Butticè; Tarik Emre Sener; Christopher Netsch; Esteban Emiliani; Rosa Pappalardo; Carlo Magno
Journal:  Cent European J Urol       Date:  2016-08-26

2.  A microbiological evaluation of level of disinfection for flexible cystoscopes protected by disposable endosheaths.

Authors:  Peter Hjorth Jørgensen; Torsten Slotsbjerg; Henrik Westh; Vicki Buitenhuis; Gregers Gautier Hermann
Journal:  BMC Urol       Date:  2013-10-07       Impact factor: 2.264

  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.