Temperature calibration formula for activated charcoal radon collectors.

Alexandre Cooper1, Thiem Ngoc Le, Takeshi Iimoto, Toshiso Kosako.   

Abstract

Radon adsorption by activated charcoal collectors such as PicoRad radon detectors is known to be largely affected by temperature and relative humidity. Quantitative models are, however, still needed for accurate radon estimation in a variable environment. Here we introduce a temperature calibration formula based on the gas adsorption theory to evaluate the radon concentration in air from the average temperature, collection time, and liquid scintillation count rate. On the basis of calibration experiments done by using the 25 m³ radon chamber available at the National Institute of Radiological Sciences in Japan, we found that the radon adsorption efficiency may vary up to a factor of two for temperatures typical of indoor conditions. We expect our results to be useful for establishing standardized protocols for optimized radon assessment in dwellings and workplaces.
Copyright © 2010 Elsevier Ltd. All rights reserved.

Mesh:

Substances:

Year:  2011        PMID: 20937546     DOI: 10.1016/j.jenvrad.2010.09.004

Source DB:  PubMed          Journal:  J Environ Radioact        ISSN: 0265-931X            Impact factor:   2.674


  1 in total

1.  New generation of highly sensitive radon detectors based on activated carbon with compensated temperature dependence.

Authors:  Dobromir Pressyanov
Journal:  Sci Rep       Date:  2022-05-19       Impact factor: 4.996