Literature DB >> 20936593

[The use of DensiProbe™ in hindfoot arthrodesis. Can fusion failure be predicted by mechanical bone strength determination?].

K Klos1, T Mückley, D Wähnert, H Zwipp, B G Gueorguiev, K Schwieger, G O Hofmann, M Windolf.   

Abstract

AIM: Bone quality is a main factor in implant fixation. After having shown promising results, we have further investigated the use of the DensiProbe™ as an intraoperative measurement tool for evaluation of calcaneal bone quality and prediction of nailed hindfoot arthrodesis failure.
METHOD: In this add-on study 19 nail arthrodeses were performed using a conventional screw plus a locked blade (n = 6) or plus a locked screw (n = 13) in the calcaneus. A specially devised tool was inserted at the fixation sites of the screws and the cancellous break-away torque was measured. The constructs were then cyclically loaded to failure.
RESULTS: We saw a wide range of BMD (1.9-185.9 mgHA/cm³, mean 102.4 mg/cm³, SD 53.5). The peak torque was 0.47-1.78 Nm (mean 0.92 Nm, SD 0.46) at the proximal screw site (PSS) and 0.24 and 1.2 Nm (mean 0.63 Nm, SD 0.37) at the distal screw site (DSS), respectively, and 0.42 and 1.52 Nm (mean 1.00 Nm, SD 0.36) in the screw plus blade group (PSS). The number of cycles correlated with peak torque (two screws group PSS: p = 0.002, r² = DSS: 0.61 p = 0.001, r² = 0.90; screw plus blade group PSS: p = 0.001, r² = 0.99). Peak torque also correlated with BMD in both groups (two screws group PSS: p = 0.01, r² = 0.71; DSS: p = 0.001; r² = 0.83; screw plus blade group PSS: 0.42 and 1.52 Nm, mean 1.00 Nm, SD 0.36).
CONCLUSION: A mechanical bone measurement tool like the DensiProbe™ seems to be suitable for predicting tibiotalocalcaneal arthrodesis failure in a biomechanical test set-up. As a restriction in clinical practice failure is multifactorial and prediction cannot be based upon these measurements only. © Georg Thieme Verlag KG Stuttgart · New York.

Entities:  

Mesh:

Year:  2010        PMID: 20936593     DOI: 10.1055/s-0030-1250106

Source DB:  PubMed          Journal:  Z Orthop Unfall        ISSN: 1864-6697            Impact factor:   0.923


  4 in total

1.  Removal of cement-augmented screws in distal femoral fractures and the effect of retained screws and cement on total knee arthroplasty: a biomechanical investigation.

Authors:  Thomas Vordemvenne; Dominic Gehweiler; Dirk Wähnert; Niklas Grüneweller; Boyko Gueorguiev
Journal:  J Orthop Traumatol       Date:  2021-02-27

2.  Local bone quality measure and construct failure prediction: a biomechanical study on distal femur fractures.

Authors:  Thomas Vordemvenne; Dirk Wähnert; Dominic Gehweiler; Ursula Styger; Boyko Gueorguiev; Christian Colcuc
Journal:  Arch Orthop Trauma Surg       Date:  2021-02-15       Impact factor: 2.928

3.  Osteoarthritis of the Distal Interphalangeal and First Carpometacarpal Joints is Associated with High Bone Mass in Women and Small Bone Size and Low Lean Mass in Men.

Authors:  Thord von Schewelov; Håkan Magnusson; Maria Cöster; Caroline Karlsson; Björn E Rosengren
Journal:  Open Orthop J       Date:  2015-08-31

4.  Bone quality assessment for total hip arthroplasty with intraoperative trabecular torque measurements.

Authors:  Matthias C M Klotz; Nicholas A Beckmann; Rudi G Bitsch; Elisabeth Seebach; Tobias Reiner; Sebastian Jäger
Journal:  J Orthop Surg Res       Date:  2014-11-13       Impact factor: 2.359

  4 in total