Literature DB >> 20926328

An open-loop controlled active lung simulator for preterm infants.

Stefano Cecchini1, Emiliano Schena, Sergio Silvestri.   

Abstract

We describe the underlying theory, design and experimental evaluation of an electromechanical analogue infant lung to simulate spontaneous breathing patterns of preterm infants. The aim of this work is to test the possibility to obtain breathing patterns of preterm infants by taking into consideration the air compressibility. Respiratory volume function represents the actuation pattern, and pulmonary pressure and flow-rate waveforms are mathematically obtained through the application of the perfect gas and adiabatic laws. The mathematical model reduces the simulation interval into a step shorter than 1 ms, allowing to consider an entire respiratory act as composed of a large number of almost instantaneous adiabatic transformations. The device consists of a spherical chamber where the air is compressed by four cylinder-pistons, moved by stepper motors, and flows through a fluid-dynamic resistance, which also works as flow-rate sensor. Specifically designed software generates the actuators motion, based on the desired ventilation parameters, without controlling the gas pneumatic parameters with a closed-loop. The system is able to simulate tidal volumes from 3 to 8 ml, breathing frequencies from 60 to 120 bpm and functional residual capacities from 25 to 80 ml. The simulated waveforms appear very close to the measured ones. Percentage differences on the tidal volume waveform vary from 7% for the tidal volume of 3 ml, down to 2.2-3.5% for tidal volumes in the range of 4-7 ml, and 1.3% for the tidal volume equal to 8 ml in the whole breathing frequency and functional residual capacity ranges. The open-loop electromechanical simulator shows that gas compressibility can be theoretically assessed in the typical pneumatic variable range of preterm infant respiratory mechanics.
Copyright © 2010 IPEM. Published by Elsevier Ltd. All rights reserved.

Entities:  

Mesh:

Year:  2011        PMID: 20926328     DOI: 10.1016/j.medengphy.2010.09.001

Source DB:  PubMed          Journal:  Med Eng Phys        ISSN: 1350-4533            Impact factor:   2.242


  2 in total

1.  An active one-lobe pulmonary simulator with compliance control for medical training in neonatal mechanical ventilation.

Authors:  Ilaria Baldoli; Selene Tognarelli; Francesca Cecchi; Rosa Teresa Scaramuzzo; Massimiliano Ciantelli; Marzia Gentile; Armando Cuttano; Cecilia Laschi; Arianna Menciassi; Antonio Boldrini; Paolo Dario
Journal:  J Clin Monit Comput       Date:  2013-10-15       Impact factor: 2.502

2.  A new infant hybrid respiratory simulator: preliminary evaluation based on clinical data.

Authors:  Barbara Stankiewicz; Krzysztof J Pałko; Marek Darowski; Krzysztof Zieliński; Maciej Kozarski
Journal:  Med Biol Eng Comput       Date:  2017-03-25       Impact factor: 2.602

  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.