Evaluation of the dosimetric parameters for 125I brachytherapy determined in prostate medium using CT images.

Takashi Hanada1, Atsunori Yorozu, Toshio Ohashi, Naoyuki Shigematsu, Koichi Maruyama.   

Abstract

In the present study, the prostate medium determined from the CT images of 149 patients was developed. The dosimetric parameters such as Λ, g(L)(r) and F(r, θ) used in TG-43U1-based calculation for an iodine-125 ((125)I) brachytherapy-source were examined using Monte Carlo code Geant4. Clinical dosimetry parameters such as the D(90) were evaluated among a subgroup of 50 randomly selected patients who had been treated with permanent brachytherapy between January 2008 and December 2008 at the Tokyo Medical Center. The results show a slight difference in the dose rate constant Λ (within 1.0%). The radial dose function g(L)(r) exhibits a prominent difference in the region over 3 cm, and this difference is maintained within 2.9% in the region close to the source. The calculated values of F(r, θ) for the prostate medium were similar to values for water (within 1%), except in the longitudinal axis. A comparison of D(90) values shows a systematic dose overestimation of 2.8 ± 0.7 Gy in water, where the distribution of the differences can be seen with a spread of 1.8 ± 0.3% compared to that in prostate medium. It was concluded that the introduction of any kind of tissue correction for the TG-43U1-based calculation was not necessary to allow for the differences in elemental compositions and densities between water and prostate medium. PACS number: 87.00.00; 87.55.dk; 87.55.K-; 87.56.B-.

Entities:  

Mesh:

Substances:

Year:  2010        PMID: 20921822     DOI: 10.1269/jrr.10027

Source DB:  PubMed          Journal:  J Radiat Res        ISSN: 0449-3060            Impact factor:   2.724


  3 in total

1.  A dosimetry method for low dose rate brachytherapy by EGS5 combined with regression to reflect source strength shortage.

Authors:  Kenichi Tanaka; Kunihiko Tateoka; Osamu Asanuma; Ken-Ichi Kamo; Kaori Sato; Hiromitsu Takeda; Masaru Takagi; Masato Hareyama; Jun Takada
Journal:  J Radiat Res       Date:  2014-01-20       Impact factor: 2.724

2.  Dosimetric effect of tissue heterogeneity for (125)I prostate implants.

Authors:  Susana Maria Oliveira; Nuno José Teixeira; Lisete Fernandes; Pedro Teles; Pedro Vaz
Journal:  Rep Pract Oncol Radiother       Date:  2014-04-16

3.  Measurement of the strength of iodine-125 seed moving at unknown speed during implantation in brachytherapy.

Authors:  Kenichi Tanaka; Satoru Endo; Kunihiko Tateoka; Osamu Asanuma; Ken-Ichi Kamo; Kaori Sato; Hiromitsu Takeda; Masaru Takagi; Masato Hareyama; Jun Takada
Journal:  J Radiat Res       Date:  2013-06-30       Impact factor: 2.724

  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.