[Analysis of image quality and dose calculation accuracy in cone beam CT acquisitions with limited projection data (half scan, half fan) with regard to usability for adaptive radiation therapy treatment planning].

Matthias Kowatsch1, Peter Winkler, Brigitte Zurl, Thomas Konrad, Jörg Schröttner.   

Abstract

UNLABELLED: Subject of this study is the question of whether cone beam CT (CBCT) images with reduced projection data are suitable for use in adaptive radiation therapy (ART) treatment planning. For this purpose image quality and dose calculation accuracy depending on imaging modality were analysed. In this context, two CBCT-methods will be indicated having reduced projection data sets: Scans acquired with 200° rotation angle in order to accelerate the CBCT process (half scan), or scans with an asymmetric cone beam and detector offset, used to enlarge the field-of-view (half fan).
METHODS: For three different CBCT-modes (On-Board-Imaging, Varian Medical Systems), two of them based on reduced projection data, and a conventional multidetector CT system, the main image quality parameters were studied. Treatment plans for two phantoms were transferred to all datasets and re-computed to analyse dose calculation accuracy. Furthermore imaging dose was measured for all modalities.
RESULTS: All three CBCT-modes showed similar results with regard to image quality. It was found, that a reduction in projection data does not necessarily involve deterioration in image quality parameters. For dose calculation based on CBCT images, a good agreement with the reference plan was found, with a maximum deviation for the mean dose in regions of interest of 1.1%. Imaging dose was found to be 2.5 cGy and 2.9 cGy for the large-FOV mode and the partial rotation mode, respectively, and 5.4 cGy for the 360°-full fan mode.
Copyright © 2010. Published by Elsevier GmbH.

Mesh:

Year:  2011        PMID: 20888201     DOI: 10.1016/j.zemedi.2010.08.002

Source DB:  PubMed          Journal:  Z Med Phys        ISSN: 0939-3889            Impact factor:   4.820


  3 in total

1.  Hounsfield units variations: impact on CT-density based conversion tables and their effects on dose distribution.

Authors:  B Zurl; R Tiefling; P Winkler; P Kindl; K S Kapp
Journal:  Strahlenther Onkol       Date:  2013-11-09       Impact factor: 3.621

2.  Comparison of low-dose, half-rotation, cone-beam CT with electronic portal imaging device for registration of fiducial markers during prostate radiotherapy.

Authors:  Ngie Min Ung; Leonard Wee; Sara Lyons Hackett; Andrew Jones; Tee Sin Lim; Christopher Stirling Harper
Journal:  J Appl Clin Med Phys       Date:  2013-07-08       Impact factor: 2.102

3.  Evaluation of 4-Hz log files and secondary Monte Carlo dose calculation as patient-specific quality assurance for VMAT prostate plans.

Authors:  Philipp Szeverinski; Matthias Kowatsch; Thomas Künzler; Marco Meinschad; Patrick Clemens; Alexander F DeVries
Journal:  J Appl Clin Med Phys       Date:  2021-06-20       Impact factor: 2.102

  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.