Literature DB >> 20883758

Velocity profiles and shear strain rate variability in the USP Dissolution Testing Apparatus 2 at different impeller agitation speeds.

Ge Bai1, Yimin Wang, Piero M Armenante.   

Abstract

The fluid velocity profiles at different locations inside a standard USP Dissolution Testing Apparatus 2 were experimentally obtained via Laser Doppler Velocimetry (LDV) at three impeller agitations speeds, namely 50rpm, 75rpm and 100rpm. The experimental results were compared with the predictions obtained with Computational Fluid Dynamics (CFD) where the κ-ω model with low Reynolds number correction was used to account for turbulence effects. In general, good agreement was found between the experimental LDV velocity measurements and the CFD simulation predictions. The non-dimensional tangential, axial and radial velocity profiles (scaled with the impeller tip speed) and the flow pattern were found to be nearly independent of the agitation speed in most regions of the vessel, implying that increasing the agitation speed generally produced a corresponding increase in the local values of the velocity. However, the velocity profiles and flow pattern in the inner core region just below the impeller, where the dissolving tablet is usually located, were found to be much less sensitive to agitation speed. In this region, the axial and radial velocities were especially low and were not significantly affected by agitation increases. This inner core region at the center of the vessel bottom persisted irrespective of agitation intensity. The CFD predictions also indicated that increasing the agitation speed resulted in a higher shear strain rate distribution along the vessel bottom, although the strain rate was always very low at the center of the vessel bottom, even when the agitation speed was increased.
Copyright © 2010 Elsevier B.V. All rights reserved.

Entities:  

Mesh:

Substances:

Year:  2010        PMID: 20883758     DOI: 10.1016/j.ijpharm.2010.09.022

Source DB:  PubMed          Journal:  Int J Pharm        ISSN: 0378-5173            Impact factor:   5.875


  5 in total

1.  Particle Image Velocimetry Evaluation of Fluid Flow Profiles in USP 4 Flow-Through Dissolution Cells.

Authors:  Hiroyuki Yoshida; Akemi Kuwana; Hiroko Shibata; Ken-Ichi Izutsu; Yukihiro Goda
Journal:  Pharm Res       Date:  2015-03-20       Impact factor: 4.200

2.  Molecular cycloencapsulation augments solubility and improves therapeutic index of brominated noscapine in prostate cancer cells.

Authors:  Jitender Madan; Bharat Baruah; Mulpuri Nagaraju; Mohamed O Abdalla; Clayton Yates; Timothy Turner; Vijay Rangari; Donald Hamelberg; Ritu Aneja
Journal:  Mol Pharm       Date:  2012-04-27       Impact factor: 4.939

3.  Impact of Select Geometric and Operational Parameters on Hydrodynamics in Dissolution Apparatus 2 (Paddle Apparatus): A Design of Experiments Analysis Based on Computational Fluid Dynamics Simulations.

Authors:  Satish Perivilli; Steven Walfish; Erika Stippler; Mark R Liddell
Journal:  Pharm Res       Date:  2022-05-16       Impact factor: 4.580

4.  An automated system for monitoring and regulating the pH of bicarbonate buffers.

Authors:  Grzegorz Garbacz; Bartosz Kołodziej; Mirko Koziolek; Werner Weitschies; Sandra Klein
Journal:  AAPS PharmSciTech       Date:  2013-03-07       Impact factor: 3.246

5.  Velocity Field Visualization in USP Dissolution Apparatus 3 Using Particle Image Velocimetry.

Authors:  Satish Perivilli; Richard Prevost; Erika Stippler
Journal:  Pharm Res       Date:  2017-04-13       Impact factor: 4.200

  5 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.